OLYMPUS ENDOTHERAPY DISPOSABLE GUIDEWIRE
Report
- Report Number
- 9680001-2010-00004
- Event Type
- Other
- Date Received
- July 28, 2010
- Date of Event
- June 1, 2010
- Report Date
- July 27, 2010
- Manufacturer
- ACCELLENT INC.
- Product Code
- OCY
- PMA / PMN Number
- K072354
- Report Source
- Manufacturer report
- Reporter Location
- GM
- Reporter Occupation
- PHYSICIAN
Narratives
THE MANUFACTURING BATCH RECORD REVIEW CONFIRMED THAT THE DEVICE MET ALL MATERIAL, ASSEMBLY AND PERFORMANCE SPECIFICATIONS AND THERE WERE NO SIGNIFICANT EVENTS OR POTENTIAL ISSUES THAT COULD RESULT IN THE DETACHMENT OF THE GUIDEWIRE TIP. THERE WERE NO OTHER COMPLAINTS NOTED FOR THIS BATCH. ANALYSIS OF THE RETURNED SAMPLE INDICATED THAT THE PRODUCT MET ALL PRODUCT DIMENSIONAL REQUIREMENTS AND SPECIFICATIONS, AND THAT THE DEVICE WAS SUBJECTED TO A FORCE CAPABLE TO MODIFY THE ORIGINAL SHAPE OF THE DEVICE AND TO FORCE THE PEBAX TO SLIP OFF THE WIRE. A NEARLY IDENTICAL FAILURE MODE HAS ALSO BEEN REPORTED BY OTHER MANUFACTURERS OF SIMILAR DEVICES THAT UTILIZE SIMILAR DESIGN FEATURES. THESE INCIDENTS HAVE INVOLVED DIFFICULTY IN PASSING THE WIRE THROUGH THE CATHETER OR STENT OR REMOVING THE WIRE. THE HEALTH HAZARD ANALYSIS AND RISK ASSESSMENT BASED ON THE INFO TO DATE INDICATED THAT THE RISK OF ADVERSE OUTCOMES ASSOCIATED WITH FAILURES OF THIS TYPE IS REMOTE AND AS LOW AS REASONABLY PRACTICAL. THE MOST PROBABLE ROOT CAUSE IS OPERATIONAL CONTEXT WHERE DEVICE PERFORMANCE WAS LIMITED DO TO ANATOMICAL/PROCEDURAL FACTORS. FURTHER ANALYSIS IS ONGOING.
DURING BILE DUCT STENT EXCHANGE, THE CUSTOMER DETECTED THE LOOSENING OF THE SHEATHING MATERIAL OF THE MOST FLEXIBLE PART OF THE GUIDE WIRE, "SO THAT THE MENTIONED MATERIAL REMAINED INSIDE THE BILE DUCT." THE USER DECIDED TO REMOVE THIS MATERIAL TOGETHER WITH THE NEXT STENT EXCHANGE, WITHIN THE NEXT 3 MONTHS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | OLYMPUS ENDOTHERAPY DISPOSABLE GUIDEWIRE | ERCP GUIDEWIRE | OCY | ACCELLENT INC. | GV230-3527S | 680892 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Required Intervention |