ZIMMER TRABECULAR METAL HUMERAL STEM
Report
- Report Number
- 1822565-2010-00534
- Event Type
- Injury
- Date Received
- July 29, 2010
- Date of Event
- June 23, 2010
- Report Date
- June 23, 2010
- Manufacturer
- ZIMMER INC
- Product Code
- HSD
- Removal / Correction Number
- 1822565-7/26/2010-006-R
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NC, US
- Reporter Occupation
- OTHER
Narratives
EVALUATION SUMMARY: THE NORMAL FIT OF THE IMPLANT TO THE BONE PREPARATION IS INTENDED TO PROVIDE AN INTERFERENCE FIT CONDITION TO PROVIDE MECHANICAL STABILITY IN THE ABSENCE OF BONE CEMENT. DEPENDING ON PT BONE QUALITY, AND THE AMOUNT OF BONE REMOVED, THE REMAINING BONE MAY NOT ALLOW THE IMPLANT TO SEAT WITH NORMAL IMPACT FORCES IN SOME CASES. WHILE A CAUSE FOR THIS SPECIFIC CLINICAL EVENT CANNOT BE ASCERTAINED FROM THE INFO PROVIDED, THE COMMON CLINICAL PRESENTATION AND THE DATE OF THE OCCURRENCE SUGGESTS THAT THE EVENT IS RELATED TO THE ISSUES FOR WHICH ZIMMER IS IMPLEMENTING CORRECTIVE ACTION. REFERENCE REMOVAL REPORT 1822565-7/26/2010-006-R FOR ADDITIONAL INFO REGARDING THE CORRECTIVE ACTION TAKEN. REVIEW OF THE DEVICE HISTORY RECORDS DID NOT FIND ANY DEVIATIONS OR ANOMALIES. IT IS NOT SUSPECTED THAT THE PRODUCT FAILED TO MEET SPECIFICATIONS. SHOULD ADDITIONAL SUBSTANTIVE INFO BE RECEIVED, THE COMPLAINT WILL BE REOPENED. ZIMMER, INC. CONSIDERS THE INVESTIGATION CLOSED.
IT IS REPORTED THAT UPON IMPLANTING THE STEM, THE SURGEON NOTICED THAT THE IMPLANT WAS NOT SEATING IN THE USUAL FASHION. HE ASKED TO RE-REAM DISTALLY AND PROXIMALLY. HE THEN TRIED TO PUT THE STEM DOWN ONE MORE TIME AND ENDED UP FRACTURING THE GREATER TUBEROSITY. THE FRACTURE WAS NOT DISPLACED SO IT WAS FIXED WITH THE PLACEMENT OF 5 HEAVY #5 SUTURES. THE SURGEON THEN ELECTED TO USE A SIZE 8MM STEM BECAUSE, THE PT WAS RHEUMATOID AND HE DID NOT WANT TO FRACTURE HER DISTALLY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ZIMMER TRABECULAR METAL HUMERAL STEM | SHOULDER PROSTHESIS | HSD | ZIMMER INC | NA | 61392762 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 46 YR | Other |