FDA Adverse Event Injury Summary report: N

UNKNOWN TEXTURED SALINE BREAST PROSTHESIS

MDR report key: 17791276 · Received September 21, 2023

Report

Report Number
1645337-2023-11244
Event Type
Injury
Date Received
September 21, 2023
Date of Event
August 15, 2023
Manufacturer
MENTOR TEXAS
Product Code
FWM
PMA / PMN Number
P990075
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
OTHER
Health Professional
N

Narratives

Additional Manufacturer Narrative · 0

ON OCTOBER 09, 2023, MENTOR RECEIVED ADDITIONAL INFORMATION REGARDING THE PATIENT'S DATE OF EVENT, AND DEVICE DATE OF EXPLANTATION. IT WAS REPORTED THAT THE PATIENT DATE OF EVENT WAS AUGUST 15, 2023. IT WAS REPORTED THAT THE PATIENT IS TO UNDERGO DEVICE EXPLANTATION ON (B)(6) 2023. SECTION D6B. EXPLANTATION DATE: (B)(6) 2023. ALL RELEVANT FIELDS HAVE BEEN UPDATED TO REFLECT THIS ADDITIONAL INFORMATION. SINCE THE DEVICE HAS NOT BEEN RETURNED FOR ANALYSIS, NO PRODUCT FAILURE ANALYSIS CAN BE CONDUCTED, AND NO DETERMINATION OF POSSIBLE CONTRIBUTING FACTORS CAN BE MADE. AS SUCH, THE INVESTIGATION WILL BE CLOSED. IF THE COMPLAINT DEVICE IS RECEIVED IN THE FUTURE, THE INVESTIGATION WILL BE REOPENED AND CONDUCTED AS APPROPRIATE. REASON FOR DEVICE EXPLANT AND/OR REOPERATION: DEFLATION. MANUFACTURER¿S REFERENCE NUMBER: (B)(4).

Additional Manufacturer Narrative · 0

ON FEBRUARY 12, 2024, MENTOR RECEIVED ADDITIONAL INFORMATION REGARDING THE DEVICE DATE OF EXPLANTATION. IT WAS REPORTED THAT THE PATIENT IS TO UNDERGO DEVICE EXPLANTATION ON (B)(6) 2024. SECTION D6B.-EXPLANTATION DATE HAS BEEN UPDATED TO (B)(6) 2024. MANUFACTURER¿S REFERENCE NUMBER: (B)(4).

Additional Manufacturer Narrative · 0

ON APRIL 04, 2024, THE MENTOR FAILURE ANALYSIS LAB HAS RECEIVED TWO DEVICES FOR EVALUATION. ON APRIL 10, 2024, MENTOR RECEIVED ADDITIONAL INFORMATION REGARDING THE PATIENT'S REPLACEMENT DEVICE. IT WAS REPORTED THAT THE PATIENT'S REPLACEMENT DEVICE WAS A 275CC MENTOR SMOOTH ROUND MODERATE PROFILE SALINE BREAST PROSTHESIS WITH SERIAL NUMBER 9934135-055. SINCE BOTH DEVICES RECEIVED CONTAIN UNKNOWN LOT NUMBERS, AND CANNOT DETERMINE WHICH IS THE IMPACTED SIDE, BOTH DEVICE SUMMARIES WILL BE PROVIDED. ON APRIL 11, 2024, THE EVALUATION FOR THE DEVICES WAS COMPLETED. DEVICE EVALUATION SUMMARY A: VISUAL ANALYSIS OF THE RETURNED SAMPLE REVEALED THAT NO DAMAGE OR ANOMALIES WERE OBSERVED ON THE TEXTURE SALINE 300CC BREAST IMPLANT. LEAK TESTING WAS PERFORMED, IN ACCORDANCE WITH MENTOR PROCEDURES, AND A TEAR WAS NOTED ON THE POSTERIOR VIEW MEASURING APPROXIMATELY 0.1 CM. MICROSCOPIC EXAMINATION WAS PERFORMED AND THE CAUSE OF THE DEFLATION COULD NOT BE IDENTIFIED. ALTHOUGH NO CONCLUSION COULD BE REACHED ON THE CAUSE OF THE REPORTED EVENT, THE INSTRUCTIONS FOR USE CONTAIN THE FOLLOWING CAUTION: CAUSES OF DEFLATION OF SALINE IMPLANTS INCLUDE BUT ARE NOT LIMITED TO THE FOLLOWING EVENTS: CREASE FOLD FAILURE, INTRAOPERATIVE OR POSTOPERATIVE TRAUMA, EXCESSIVE STRESSES OR MANIPULATIONS AS MAY OCCUR DURING NORMAL, DAILY ROUTINES INCLUDING CUSTOMARY AND PURPOSEFUL TRAUMA SUCH AS THAT WHICH CAN OCCUR DURING VIGOROUS EXERCISE, ATHLETICS, ETC. AS PART OF MENTOR¿S QUALITY PROCESS, ALL DEVICES ARE MANUFACTURED, INSPECTED, AND RELEASED TO APPROVED SPECIFICATIONS. NO CORRECTIVE AND PREVENTIVE ACTION (CAPA) IS REQUIRED NOW. ADDITIONAL COMPLAINT INFORMATION MONITORING FOR POTENTIAL SAFETY SIGNALS IS CONDUCTED THROUGH COMPLAINT TRENDING AS PART OF POST-MARKET SURVEILLANCE. DEVICE EVALUATION SUMMARY B: VISUAL ANALYSIS OF THE RETURNED SAMPLE REVEALED THAT THE TEXTURED SALINE 300CC BREAST IMPLANT WAS FOUND TO HAVE A TEAR ON THE ANTERIOR VIEW EXTENDING TO THE POSTERIOR VIEW MEASURING APPROXIMATELY 4.5 CM. MICROSCOPIC EXAMINATION WAS PERFORMED AND THE CAUSE OF THE DEFLATION COULD NOT BE IDENTIFIED. ALTHOUGH NO CONCLUSION COULD BE REACHED ON THE CAUSE OF THE REPORTED EVENT, THE INSTRUCTIONS FOR USE CONTAIN THE FOLLOWING CAUTION: CAUSES OF DEFLATION OF SALINE IMPLANTS INCLUDE BUT ARE NOT LIMITED TO THE FOLLOWING EVENTS: CREASE FOLD FAILURE, INTRAOPERATIVE OR POSTOPERATIVE TRAUMA, EXCESSIVE STRESSES OR MANIPULATIONS AS MAY OCCUR DURING NORMAL, DAILY ROUTINES INCLUDING CUSTOMARY AND PURPOSEFUL TRAUMA SUCH AS THAT WHICH CAN OCCUR DURING VIGOROUS EXERCISE, ATHLETICS, ETC. AS PART OF MENTOR¿S QUALITY PROCESS, ALL DEVICES ARE MANUFACTURED, INSPECTED, AND RELEASED TO APPROVED SPECIFICATIONS. NO CORRECTIVE AND PREVENTIVE ACTION (CAPA) IS REQUIRED NOW. ADDITIONAL COMPLAINT INFORMATION MONITORING FOR POTENTIAL SAFETY SIGNALS IS CONDUCTED THROUGH COMPLAINT TRENDING AS PART OF POST-MARKET SURVEILLANCE. ON APRIL 12, 2024, MENTOR RECEIVED CLARIFICATION REGARDING THE IMPACTED SIDES. IT WAS REPORTED THAT THE PATIENT'S LEFT SIDE WAS INTACT DURING SURGERY AND ONLY THE RIGHT SIDE WAS FOUND TO BE RUPTURED. THE LEFT SIDE WAS RUPTURED BY THE SURGEON ON PURPOSE TO MEASURE THE AMOUNT OF SALINE. MANUFACTURER¿S REFERENCE NUMBER: (B)(4).

Additional Manufacturer Narrative · 0

AT THE TIME OF THIS REPORT, MENTOR HAS RECEIVED NO INFORMATION REGARDING EXPLANTATION OR AN EXPECTED EXPLANTATION DATE. IT IS UNKNOWN AT THIS TIME IF THE DEVICE WILL BE MADE AVAILABLE FOR RETURN. AS A RESULT, NO PRODUCT FAILURE ANALYSIS CAN BE CONDUCTED, AND NO DETERMINATION OF POSSIBLE CONTRIBUTING FACTORS CAN BE MADE. AS SUCH, THE INVESTIGATION WILL BE CLOSED. IF THE COMPLAINT DEVICE IS RECEIVED IN THE FUTURE, THE INVESTIGATION WILL BE REOPENED AND CONDUCTED AS APPROPRIATE. SINCE NO LOT NUMBER WAS PROVIDED, NO MANUFACTURING RECORD EVALUATION REVIEW COULD BE PERFORMED. REASON FOR DEVICE EXPLANT AND/OR REOPERATION: NO INFORMATION REGARDING EXPLANTATION OR AN EXPECTED EXPLANTATION DATE HAS BEEN RECEIVED. MENTOR IS SUBMITTING THIS REPORT PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH MENTOR HAS NOT BEEN ABLE TO INVESTIGATE OR VERIFY PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY FDA, MENTOR, OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE DEVICE, MENTOR, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. IF CERTAIN INFORMATION IS UNKNOWN, NOT AVAILABLE OR DOES NOT APPLY, THE SECTION/FIELD OF THE FORM IS LEFT BLANK. MANUFACTURER¿S REFERENCE NUMBER: (B)(4).

Description of Event or Problem · 0

IT WAS REPORTED THAT A 65-YEAR-OLD CAUCASIAN FEMALE PATIENT UNDERWENT A BREAST AUGMENTATION WITH AN UNSPECIFIED TEXTURED SALINE BREAST PROSTHESIS AND EXPERIENCED A DEFLATION ON THE RIGHT SIDE POST-OPERATIVELY. THE PATIENT MENTIONED IT HAD A SLOW LEAK AND WAS DISTORTED. AT THE TIME OF THIS REPORT, MENTOR HAS RECEIVED NO INFORMATION REGARDING EXPLANTATION OR AN EXPECTED EXPLANTATION DATE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1595129 UNKNOWN TEXTURED SALINE BREAST PROSTHESIS PROSTHESIS, BREAST, INFLATABLE, INTERNAL, SALINE FWM MENTOR TEXAS

Patients

Seq Age Sex Outcome Treatment
1 65 YR Female Required Intervention| O