FDA Adverse Event Malfunction Summary report: N

MIDMARK ULTRA TRIM DENTAL CHAIR

MDR report key: 17790004 · Received September 21, 2023

Report

Report Number
1523530-2023-00005
Event Type
Malfunction
Date Received
September 21, 2023
Date of Event
August 21, 2023
Report Date
September 21, 2023
Manufacturer
MIDMARK CORPORATION
Product Code
NRU
UDI-DI
00841709108976
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
003

Narratives

Description of Event or Problem · 0

THIS REPORT SUMMARIZES AN EVENT THAT WAS REPORTED TO MIDMARK ON (B)(6), 2023 IN WHICH IT WAS STATED THAT THE BACK OF A DENTAL CHAIR DETACHED FROM THE BASE WITH A PATIENT IN THE CHAIR. NO INJURY WAS REPORTED. PER SERVICE NOTE, THE SHOULDER BOLTS THAT HOLD THE BACKREST, AT THE SIDE OF THE CHAIR, BEHIND THE TOUCH PAD CONTROLS APPEARED LOOSE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1584762 MIDMARK ULTRA TRIM DENTAL CHAIR ULTRA TRIM DENTAL CHAIR NRU MIDMARK CORPORATION 153758-003 00841709108976

Patients

Seq Age Sex Outcome Treatment
1 Unknown Other