FDA Adverse Event Malfunction Summary report: N

AUTOLOG IQ

MDR report key: 17788265 · Received September 21, 2023

Report

Report Number
2184009-2023-00949
Event Type
Malfunction
Date Received
September 21, 2023
Date of Event
July 11, 2023
Report Date
November 8, 2023
Manufacturer
PERFUSION SYSTEMS
Product Code
CAC
UDI-DI
00643169982864
PMA / PMN Number
K181954
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CT, US
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

DEVICE EVALUATION:THE REPORTED ISSUE THAT THE INSTRUMENT WAS NOT FILLING THE BOWL, AND THE BLOOD PRODUCT IS BEING DUMPED DIRECTLY INTO THE WASTE BAG WAS VERIFIED DURING SERVICE. THE SERVICE TECHNICIAN FOUND THE LEVEL SENSOR OPTICS OUT OF CALIBRATION AND WAS UNABLE TO CALIBRATE. THE SERVICE TECHNICIAN REPLACED THE OPTICS EMITTER AND DETECTOR SENSOR AND THE CABLES X2 . WHEN THE SERVICE TECHNICIAN REMOVED THE SENSORS THERE WAS EVIDENCE OF BLOOD INTRUSION ON THE SENSORS. THE SERVICE TECHNICIAN COMPLETED THE HLT AND ALL TESTS PASSED. POST REPAIR TESTING WAS PERFORMED PER SPECIFICATIONS. NOTE:THE INSTRUMENT WAS NOT RETURNED TO MEDTRONIC FACILITY BUT WAS SERVICED BY FIELD SERVICE TECHNICIAN. MEDTRONIC IS SUBMITTING THIS REPORT TO COMPLY WITH FDA REPORTING REGULATIONS UNDER 21 CFR PARTS 4 AND 803. THIS REPORT IS BASED UPON INFORMATION OBTAINED BY MEDTRONIC, WHICH THE COMPANY MAY NOT HAVE BEEN ABLE TO FULLY INVESTIGATE OR VERIFY PRIOR TO THE DATE THE REPORT WAS REQUIRED BY THE FDA. MEDTRONIC HAS MADE REASONABLE EFFORTS TO OBTAIN MORE COMPLETE INFORMATION AND HAS PROVIDED AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. IN PARTICULAR, THIS REPORT DOES NOT CONSTITUTE AN ADMISSION BY ANYONE THAT THE PRODUCT DESCRIBED IN THIS REPORT HAS ANY ¿DEFECTS¿ OR HAS ¿MALFUNCTIONED¿. THESE WORDS ARE INCLUDED IN THE FDA 3500A FORM AND ARE FIXED ITEMS FOR SELECTION CREATED BY THE FDA TO CATEGORIZE THE TYPE OF EVENT SOLELY FOR THE PURPOSE OF REGULATORY REPORTING. MEDTRONIC OBJECTS TO THE USE OF THESE WORDS AND OTHERS LIKE THEM BECAUSE OF THE LACK OF DEFINITION AND THE CONNOTATIONS IMPLIED BY THESE TERMS. THIS STATEMENT SHOULD BE INCLUDED WITH ANY INFORMATION OR REPORT DISCLOSED TO THE PUBLIC UNDER THE FREEDOM OF INFORMATION ACT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. A GOOD FAITH EFFORT WILL BE MADE TO OBTAIN THE APPLICABLE INFORMATION RELEVANT TO THE REPORT. IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Additional Manufacturer Narrative · 0

CONCLUSION: REPORTED EVENT WAS CONFIRMED. DUE TO THE AVAILABLE INFORMATION AND THE AGE OF THE INSTRUMENT THE MOST LIKELY CAUSE IS NORMAL COMPONENT WEAR OUT. ALL ASSOCIATED RISKS ARE LOW AND NO DEVICE RELATED PATIENT/CLINICAL SAFETY ISSUES WERE REPORTED. TRENDS FOR ISSUES WITH THIS PRODUCT ARE REVIEWED AT QUARTERLY QUALITY MEETINGS. MEDTRONIC IS SUBMITTING THIS REPORT TO COMPLY WITH FDA REPORTING REGULATIONS UNDER 21 CFR PARTS 4 AND 803. THIS REPORT IS BASED UPON INFORMATION OBTAINED BY MEDTRONIC, WHICH THE COMPANY MAY NOT HAVE BEEN ABLE TO FULLY INVESTIGATE OR VERIFY PRIOR TO THE DATE THE REPORT WAS REQUIRED BY THE FDA. MEDTRONIC HAS MADE REASONABLE EFFORTS TO OBTAIN MORE COMPLETE INFORMATION AND HAS PROVIDED AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. IN PARTICULAR, THIS REPORT DOES NOT CONSTITUTE AN ADMISSION BY ANYONE THAT THE PRODUCT DESCRIBED IN THIS REPORT HAS ANY ¿DEFECTS¿ OR HAS ¿MALFUNCTIONED¿. THESE WORDS ARE INCLUDED IN THE FDA 3500A FORM AND ARE FIXED ITEMS FOR SELECTION CREATED BY THE FDA TO CATEGORIZE THE TYPE OF EVENT SOLELY FOR THE PURPOSE OF REGULATORY REPORTING. MEDTRONIC OBJECTS TO THE USE OF THESE WORDS AND OTHERS LIKE THEM BECAUSE OF THE LACK OF DEFINITION AND THE CONNOTATIONS IMPLIED BY THESE TERMS. THIS STATEMENT SHOULD BE INCLUDED WITH ANY INFORMATION OR REPORT DISCLOSED TO THE PUBLIC UNDER THE FREEDOM OF INFORMATION ACT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. A GOOD FAITH EFFORT WILL BE MADE TO OBTAIN THE APPLICABLE INFORMATION RELEVANT TO THE REPORT. IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Description of Event or Problem · 0

MEDTRONIC RECEIVED INFORMATION THAT DURING USE OF AN AUTOLOG IQ INSTRUMENT, IT WAS REPORTED THAT THE INSTRUMENT WAS NOT FILLING THE BOWL, AND THE BLOOD PRODUCT IS BEING DUMPED DIRECTLY INTO THE WASTE BAG. USE OF INSTRUMENT WAS UNSPECIFIED. THERE WAS NO ADVERSE PATIENT EFFECT ASSOCIATED WITH THIS EVENT. MEDTRONIC RECEIVED ADDITIONAL INFORMATION THAT THE BLOOD SENT TO THE WASTE BAG WAS DISCARDED AND A TRANSFUSION WAS NOT REQUIRED AS A RESULT OF THIS ISSUE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1817781 AUTOLOG IQ APPARATUS, AUTOTRANSFUSION CAC PERFUSION SYSTEMS ATLGIQ 00643169982864

Patients

Seq Age Sex Outcome Treatment
1 Unknown