HAMILTON MEDICAL AG
Report
- Report Number
- 3001421318-2023-18886
- Event Type
- Malfunction
- Date Received
- September 21, 2023
- Date of Event
- November 15, 2021
- Report Date
- September 20, 2023
- Manufacturer
- HAMILTON MEDICAL AG
- Product Code
- CBK
- UDI-DI
- 00730002856789
- PMA / PMN Number
- K201306
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
SINCE THE COMPLAINT IN QUESTION WAS SUBMITTED TO HAMILTON MEDICAL AG ALMOST 2 YEARS AGO, NO ATTEMPTS WILL BE PERFORMED TO OBTAIN ADDITIONAL INFORMATION. NO FURTHER INVESTIGATION OR CORRECTION WILL BE PERFORMED EXCEPT THOSE MENTIONED ABOVE. IN FUTURE HAMILTON MEDICAL AG WILL REPORT AN EVENT SIMILAR TO THIS ISSUE AS IT WILL BE DEEMED A REPORTABLE EVENT. THE ALLEGATION IN THIS COMPLAINT WAS CONFIRMED TO BE A COMPLAINT. WITH THIS INVESTIGATION IT HAS BEEN CONFIRMED THAT THE DEVICE FAILED TO MEET ITS SPECIFICATIONS AT THE TIME OF THE EVENT WHILE THE VENTILATOR WAS NOT IN USE. THE ROOT CAUSE WAS DETERMINED AS PER THE RGA#75414 TO BE A FAILURE TRACED TO A FAULTY PRESSURE SENSOR BOARD. IN CONSEQUENCE THE PRESSURE SENSOR ASSEMBLY WAS REPLACED, AND NO OTHER FAILURES WERE REGISTERED FOR THIS MACHINE. THERE WAS NO PATIENT OR USER HARM.
SENSOR FAILURE MODE VENTILATION ENDED 134/2021-11-14 16:18:27/ALARM /ALARM SILENCE ON 135/2021-11-14 16:18:27/!! /INSPIRATORY VOLUME LIMITATION CONDITION REMOVED 136/2021-11-14 16:18:27/!!! /WRONG FLOW SENSOR CONDITION REMOVED 137/2021-11-14 16:18:27/!!! /CHECK FLOW SENSOR TUBING CONDITION REMOVED 138/2021-11-14 16:18:27/!!! /DISCONNECTION ON VENTILATOR SIDE 139/2021-11-14 16:18:26/!! /LOSS OF PEEP CONDITION REMOVED 140/2021-11-14 16:18:26/!!! /LOW PRESSURE CONDITION REMOVED 141/2021-11-14 16:18:24/!!! /WRONG FLOW SENSOR.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1817737 | HAMILTON MEDICAL AG | HAMILTON-C3 | CBK | HAMILTON MEDICAL AG | HAMILTON-C3 | 00730002856789 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown |