BD SAF-T-INTIMA¿ SAFETY SYSTEM WITH Y ADAPTER
Report
- Report Number
- 3014704491-2023-00573
- Event Type
- Malfunction
- Date Received
- September 20, 2023
- Date of Event
- August 14, 2023
- Report Date
- October 13, 2023
- Manufacturer
- BD (SUZHOU)
- Product Code
- FOZ
- UDI-DI
- 00382903830336
- PMA / PMN Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CH
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
H6: INVESTIGATION SUMMARY A DEVICE HISTORY REVIEW WAS CONDUCTED FOR LOT NUMBER 3080068. OUR RECORDS SHOW THAT THIS IS THE ONLY INSTANCE OF THIS ISSUE OCCURRING IN THIS PRODUCTION BATCH. ACCORDING TO THE SAMPLING PLAN APPLIED FOR PRODUCT PERFORMANCE, THIS LOT WAS ACCEPTED AND RELEASED WITHOUT DEFECTS BEING NOTED DURING THE FINAL ASSEMBLY OR VISUAL INSPECTIONS. A SAMPLE COULD NOT BE OBTAINED FOR EVALUATION AND TESTING; IN LIEU OF THE AFFECTED DEVICE, FUNCTIONAL TESTING WAS PERFORMED ON RETENTION SAMPLES FOR THIS LOT, THE RESULTS OF THESE SHOW THAT THE TESTED UNITS PERFORMED WITHIN PRODUCT SPECIFICATIONS. UNFORTUNATELY WITHOUT THE ABILITY TO INVESTIGATE THE AFFECTED UNIT OUR QUALITY ENGINEERS WERE UNABLE TO DETERMINE THE ROOT CAUSE FOR THIS COMPLAINT. H3 OTHER TEXT : SEE H10.
H.3. A DEVICE EVALUATION AND/OR DEVICE HISTORY REVIEW IS ANTICIPATED BUT IS NOT COMPLETE. UPON COMPLETION, A SUPPLEMENTAL REPORT WILL BE FILED.
IT WAS REPORTED WHILE USING BD SAF-T-INTIMA¿ SAFETY SYSTEM WITH Y ADAPTER THE PRN WAS LOOSE. THERE WAS NO REPORT OF PATIENT IMPACT. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER, TRANSLATED FROM CHINESE TO ENGLISH: PREPARE THE PATIENT FOR PLACEMENT OF THE INDWELLING NEEDLE. THE INDWELLING NEEDLE IS INTACT, AND THE HEPARIN CAP FALLS OFF AFTER OPENING THE OUTER PACKAGE.
IT WAS REPORTED WHILE USING BD SAF-T-INTIMA¿ SAFETY SYSTEM WITH Y ADAPTER THE PRN WAS LOOSE. THERE WAS NO REPORT OF PATIENT IMPACT. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER, TRANSLATED FROM CHINESE TO ENGLISH: PREPARE THE PATIENT FOR PLACEMENT OF THE INDWELLING NEEDLE. THE INDWELLING NEEDLE IS INTACT, AND THE HEPARIN CAP FALLS OFF AFTER OPENING THE OUTER PACKAGE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1585821 | BD SAF-T-INTIMA¿ SAFETY SYSTEM WITH Y ADAPTER | INTRAVASCULAR CATHETER | FOZ | BD (SUZHOU) | 3080068 | 00382903830336 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown |