FDA Adverse Event Malfunction Summary report: N

FREESTYLE FREEDOM LITE

MDR report key: 1778164 · Received July 30, 2010

Report

Report Number
2954323-2010-01031
Event Type
Malfunction
Date Received
July 30, 2010
Date of Event
July 8, 2010
Report Date
August 24, 2010
Product Code
NBW
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE PRODUCT HAS BEEN REQUESTED BACK FOR AN INVESTIGATION. A FOLLOW-UP REPORT WILL BE FILED ONCE INVESTIGATION RESULTS ARE AVAILABLE.

Additional Manufacturer Narrative · 1

THIS IS A FINAL REPORT. THE CUSTOMER RETURNED METER (B)(4). THE METER HAS BEEN TESTED WITH RETURNED STRIP LOT 1011944. THE COMPLAINT WAS NOT CONFIRMED AND ALL RESULTS WERE WITHIN THE RANGE SPECIFICATION DURING CONTROL SOLUTION TESTING. IN ADDITION, THE READINGS REPORTED WERE FOUND IN THE INTERNAL MEMORY LOG OF THE METER AND WERE TAKEN IN 10 MINUTES.

Description of Event or Problem · 1

CUSTOMER REPORTED RECEIVING ERRATIC READINGS ON THEIR ADC BLOOD GLUCOSE METER. CUSTOMER REPORTED RECEIVING READINGS OF 24 MG/DL AND 109 MG/DL WITHIN 10 MINUTES. THE RESULTS WHEN PLOTTED ON A PARKES ERROR GRID FELL INTO THE "C" ZONE SHOWING THE DIFFERENCE IN VALUES TO BE CLINICALLY SIGNIFICANT. THERE WAS NO REPORT OF DEATH, SERIOUS INJURY, OR MISTREATMENT ASSOCIATED WITH THIS EVENT.

Description of Event or Problem · 1

IT WAS REPORTED THAT INSULIN LEAKED FROM THE RESERVOIR. NOTHING FURTHER WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 FREESTYLE FREEDOM LITE BLOOD GLUCOSE MONITORING SYSTEM NBW 1011944

Patients

Seq Age Sex Outcome Treatment
1