FDA Adverse Event Malfunction Summary report: N

ZOLL IVTM THERMOGARD XP

MDR report key: 17781458 · Received September 20, 2023

Report

Report Number
3010617000-2023-00859
Event Type
Malfunction
Date Received
September 20, 2023
Date of Event
August 25, 2023
Report Date
September 20, 2023
Manufacturer
ZOLL CIRCULATION
Product Code
NCX
UDI-DI
00849111075022
PMA / PMN Number
K072234
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

DURING DEVICE INSTALLATION AT THE CUSTOMER SITE, ZOLL DETECTED LOOSE ITEMS IN THE DISPLAY HEAD AND COULD HEAR A RATTLING SOUND FROM THE DISPLAY HEAD OF THE THERMOGARD XP IVTM SYSTEM SN (B)(6). THE DISPLAY HEAD (LOT #182409) WAS RETURNED TO THE ZOLL GERMANY SERVICE DEPOT, AND UPON INSPECTION, NOTED FOUR SCREWS OF 3 BUTTON BOARDS, INCLUDING WASHERS, WERE LOOSE INSIDE THE DISPLAY. THE DISPLAY HEAD WAS REPLACED TO ADDRESS THE OBSERVED ISSUES. NO OTHER PHYSICAL DAMAGES WERE NOTED. UPON FIXING THE SCREWS AND WASHERS, THE RETURNED DISPLAY HEAD (LOT #182409) PASSED ALL THE TESTING AND WAS RETURNED TO ZOLL SERVICE STOCK FOR FUTURE USE. UPON REPLACING THE DISPLAY HEAD, THE THERMOGARD SYSTEM WAS INSTALLED AT THE CUSTOMER SITE, AND THE SYSTEM PASSED ALL THE FUNCTIONAL TESTS WITH NO ISSUES.

Description of Event or Problem · 0

DURING UNPACKING OF THE THERMOGARD XP IVTM SYSTEM SN (B)(6) FOR INSTALLATION AT THE CUSTOMER SITE, THE ZOLL TERRITORY MANAGER DETECTED LOOSE ITEMS IN THE DISPLAY HEAD AND COULD HEAR A RATTLING SOUND FROM THE DISPLAY HEAD. ALSO, A WASHER FELL OUT OF THE DISPLAY HEAD. NO PATIENT INVOLVEMENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1614503 ZOLL IVTM THERMOGARD XP THERMAL REGULATING SYSTEM NCX ZOLL CIRCULATION 8700-0650 00849111075022

Patients

Seq Age Sex Outcome Treatment
1 Unknown