FDA Adverse Event Injury Summary report: N

MECTACER BIOLOX FORTE CERAMIC BALL HEAD 12/14 Ø 28 SIZE M 0 (USA)

MDR report key: 17779920 · Received September 20, 2023

Report

Report Number
3005180920-2023-00728
Event Type
Injury
Date Received
September 20, 2023
Date of Event
August 28, 2023
Report Date
September 20, 2023
Manufacturer
MEDACTA INTERNATIONAL SA
Product Code
LZO
PMA / PMN Number
K073337
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
MO, US
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

BATCH REVIEW PERFORMED ON 30 AUGUST 2023: LOT 092211: 50 ITEMS MANUFACTURED AND RELEASED ON 10-SEP-2009. EXPIRATION DATE: 2014-AUG-31. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, 45 ITEMS OF THE SAME LOT HAVE BEEN SOLD WITH NO SIMILAR REPORTED EVENT DURING THE PERIOD OF REVIEW. ADDITIONAL DEVICES INVOLVED: BATCH REVIEWS PERFORMED ON 30 AUGUST 2023: LINER: VERSAFITCUP DM 01.26.2850M DOUBLE MOBILITY LINER Ø 50/28 (K083116) LOT 092005: 59 ITEMS MANUFACTURED AND RELEASED ON 14-SEP-2009. EXPIRATION DATE: 2014-JUL-31. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, ALL ITEMS OF THE SAME LOT HAVE BEEN SOLD WITH NO SIMILAR REPORTED EVENT DURING THE PERIOD OF REVIEW. CUP: VERSAFITCUP 01.26.50MB ACETABULAR SHELL CC Ø 50 (K083116) LOT 090908: 40 ITEMS MANUFACTURED AND RELEASED ON 13-JUL-2009. EXPIRATION DATE: 2014-APR-30. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, ALL ITEMS OF THE SAME LOT HAVE BEEN SOLD WITH NO SIMILAR REPORTED EVENT DURING THE PERIOD OF REVIEW. STEM: QUADRA-H 01.12.23SN CEMENTLESS, HA COATED STD STEM SIZE 3, SHORT NECK (K082792) LOT 090823: 45 ITEMS MANUFACTURED AND RELEASED ON 20-MAY-2009. EXPIRATION DATE: 2014-APR-30. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, ALL ITEMS OF THE SAME LOT HAVE BEEN SOLD WITH NO SIMILAR REPORTED EVENT DURING THE PERIOD OF REVIEW.

Description of Event or Problem · 0

ABOUT 13 YEARS AND 7 MONTHS AFTER THE PRIMARY SURGERY, THE PATIENT WAS REVISED DUE TO DISCOMFORT. X-RAYS INDICATED SOME RADIOLUCENCY AROUND THE STEM AND CUP. WHEN THE SURGEON WENT IN TO REVISE, IT WAS NOTICED THAT THE CUP AND STEM WERE WELL FIXATED, BUT THE LINER AND HEAD HAD SOME WEAR AND WAS STIFF. THE SURGEON REVISED THE HEAD AND LINER AND THE SURGERY WAS COMPLETED SUCCESSFULLY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
83414 MECTACER BIOLOX FORTE CERAMIC BALL HEAD 12/14 Ø 28 SIZE M 0 (USA) HIP CERAMIC BALL HEAD LZO MEDACTA INTERNATIONAL SA 092211

Patients

Seq Age Sex Outcome Treatment
1 67 YR Female Required Intervention