REMSTAR AUTO A-FLEX
Report
- Report Number
- 2518422-2023-23762
- Event Type
- Malfunction
- Date Received
- September 20, 2023
- Date of Event
- September 7, 2023
- Report Date
- January 30, 2025
- Manufacturer
- RESPIRONICS, INC.
- Product Code
- BZD
- PMA / PMN Number
- K131982
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- DA
- Reporter Occupation
- 003
Narratives
H3 OTHER TEXT: DEVICE NOT RETURNED TO THE MANUFACTURER.
1) THE DEVICE WAS RECEIVED BY THE MANUFACTURER'S PRODUCT INVESTIGATION LABORATORY. THE DEVICE HAS BEEN EVALUATED FOR LAB INVESTIGATION. DURING THE EXTERNAL AND INTERNAL EVALUATION OF THE DEVICE, IT IS OBSERVED THAT THERE IS NO EVIDENCE OF SOUND ABATEMENT FOAM DEGRADATION/BREAKDOWN IN THE DEVICE. THIS INVESTIGATION WAS PERFORMED AS OUTLINED IN ER 2244873 (V01). BECAUSE THE DEVICE RETURNED FROM SERVICE MISSING THE SENSOR SEAL, PIL HAD TO REASSEMBLE THE DEVICE WITH PIL SUPPLIED PARTS TO VERIFY OPERATION. THE ISSUE IDENTIFIED DURING DISASSEMBLY PROCESS WITH BURNT HUMIDIFIER BASE CABLE. PIL USED A KNOWN GOOD PIL SUPPLIED HUMIDIFIER (ETF 3100682-021) ON BASE DEVICE AND VERIFIED THE HEATER PLATE DID NOT HEAT. PIL USED A PIL SUPPLIED KNOWN GOOD HEATED TUBE ON THE KNOWN GOOD PIL SUPPLIED HUMIDIFIER (ETF 3100682-021) AND VERIFIED TUBE DID NOT HEAT. HUMIDIFICATION INOPERATIVE POSSIBLY BECAUSE OF THE J9 THERMAL EVENT INV0636. PIL ALSO OBSERVED THE FOLLOWING FROM AN EXTERNAL AND INTERNAL INSPECTION: AIR INLET FILTER MISSING. UI (USER INTERFACE) KNOB BROKEN. THE AIR OUTLET PORT HAD TAN DUST-LIKE CONTAMINATION IN IT. POTENTIAL MINERAL DEPOSITS IN OUTLET PORT INDICATE POSSIBLE WATER INGRESS. AIR INLET HAD TAN DUST-LIKE CONTAMINATION IN IT. PCA (PRINTED CIRCUIT ASSEMBLY) HAD DUST-LIKE CONTAMINATION ALL OVER THE TOP OF IT. ADDITIONALLY, DUST-LIKE CONTAMINATION FOUND ON THE TOP OF THE BLOWER BOX LID AND SURROUNDING AREA, ON THE TOP OF THE BLOWER MOTOR AND INSIDE OF THE BLOWER BOX. POTENTIAL MINERAL DEPOSITS IN BLOWER BOX AND BLOWER MOTOR FOUND WHICH INDICATE POSSIBLE WATER INGRESS. BOTTOM ENCLOSURE HAD DUST-LIKE CONTAMINATION IN IT. THE SOUND ABATEMENT FOAM WAS INTACT AND HAD SIGNIFICANT DUST-LIKE CONTAMINATION ON TOP OF IT. 2) IN THE PREVIOUS FILE THE REPORT WAS SUBMITTED FOR INITIAL WHICH IS UPDATED TO FINAL IN THIS REPORT. ADDITIONALLY, EVALUATION METHOD CODE GRID, EVALUATION RESULTS CODE GRID, COMPONENT CODE GRID AND CONCLUSION CODE GRID WERE ALSO UPDATED IN THIS REPORT.
THE MANUFACTURER RECEIVED INFORMATION ALLEGING A "BURNT HUMIDIFIER-BASED CABLE" LOCATED ON THE DEVICE DURING THE DISASSEMBLY PROCESS. THERE IS NO ALLEGATION OF SERIOUS OR PERMANENT HARM OR INJURY. NO MEDICAL INTERVENTION WAS REQUIRED BY THE PATIENT. THE MANUFACTURER'S INVESTIGATION IS ONGOING. A FOLLOW-UP REPORT WILL BE SUBMITTED WHEN THE MANUFACTURER'S INVESTIGATION IS COMPLETE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1615285 | REMSTAR AUTO A-FLEX | VENTILATOR, NON-CONTINUOUS (RESPIRATOR) | BZD | RESPIRONICS, INC. | IN561S |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown |