FDA Adverse Event
Death
Summary report: N
BD THREADED LOCK CANNULA
MDR report key: 17777
·
Received October 13, 1994
Report
- Report Number
- 17777
- Event Type
- Death
- Date Received
- October 13, 1994
- Date of Event
- October 4, 1994
- Report Date
- October 12, 1994
- Manufacturer
- BECTON DICKINSON & CO.
- Product Code
- FPA
- Product Problem
- Yes
- Report Source
- User Facility report
- Reporter Location
- CA, US
- Reporter Occupation
- OTHER
Narratives
Description of Event or Problem · 1
THERE RECENTLY HAVE BEEN THREE REPORTED INCIDENTS OF AN EXTENSION TUBING BECOMING DISCONNECTED FROM A THREADED-LOCK CANNULA. THE FIRST INSTANCE ON TUESDAY/WEDNESDAY, 10/4/94 RESULTED IN A PT'S DEATH. THE OTHER TWO INSTANCES DID NOT RESULT IN PT-RELATED PROBLEMS. THE SECOND WAS ON 10/6/94 AND THE THIRD WAS OVER THE WEEKEND OF 10/8-9/94. THE LIP ON THE CONNECTOR DOES NOT APPEAR TO SECURELY ENGAGE THE RIDGES ON THE LUER LOCK. EVEN WHEN THE CONNECTION SEEMS SECURE AND DOES NOT LEAK, IT CAN SUBSEQUENTLY DISCONNECT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | BD THREADED LOCK CANNULA | CANNULA | FPA | BECTON DICKINSON & CO. |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 36 MO | Death |