TELESCOPE, 10 MM, 30°, HD, QUICK LOCK, AUTOCLAVABLE
Report
- Report Number
- 3003724334-2023-00038
- Event Type
- Malfunction
- Date Received
- September 19, 2023
- Report Date
- July 1, 2024
- Manufacturer
- OLYMPUS WINTER & IBE GMBH¿
- Product Code
- EWY
- UDI-DI
- 04042761052764
- PMA / PMN Number
- K912362
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- HK
- Reporter Occupation
- OTHER
- Health Professional
- N
Narratives
THE DEVICE WAS RETURNED TO OLYMPUS FOR EVALUATION AND THE CUSTOMER'S ALLEGATION WAS ABLE TO BE CONFIRMED. DURING THE DEVICE EVALUATION IT WAS OBSERVED THAT THE LIGHT GUIDE CONNECTOR DISLODGED AND WAS STUCK INSIDE OF THE LIGHT GUIDE CABLE. IT WAS ALSO OBSERVED THAT THERE WAS LIGHT GUIDE BREAKAGE. LASTLY, IT WAS OBSERVED THAT THERE WERE DENTS ON THE OUTER TUBE. THE INVESTIGATION IS ONGOING. A SUPPLEMENTAL REPORT WILL BE SUBMITTED UPON COMPLETION OF THE INVESTIGATION. ADDITIONAL PMA/510(K) NUMBERS: K923982/ K950076.
THIS REPORT IS BEING SUPPLEMENTED TO PROVIDE ADDITIONAL INFORMATION BASED ON THE LEGAL MANUFACTURER'S FINAL INVESTIGATION. A REVIEW OF THE DEVICE HISTORY RECORD FOUND NO DEVIATIONS THAT COULD HAVE CAUSED OR CONTRIBUTED TO THE REPORTED ISSUE. BASED ON THE RESULTS OF THE INVESTIGATION, THE ROOT CAUSE OF THE EVENT WAS UNABLE TO BE IDENTIFIED. THE EVENT LIKELY OCCURRED DUE TO USER ERROR, IMPROPER HANDLING AS WELL AS APPLICATION OF EXCESSIVE FORCE. OLYMPUS WILL CONTINUE TO MONITOR FIELD PERFORMANCE FOR THIS DEVICE.
THIS REPORT IS BEING SUBMITTED TO CORRECT THE LEGAL MANUFACTURER¿S CONTACT INFORMATION AND FACILITY REGISTRATION NUMBER. THE FACILITY REGISTRATION NUMBER IS 9610773.
THE CUSTOMER REPORTED TO OLYMPUS THE TELESCOPE, 10 MM, 30°, HD, QUICK LOCK, AUTOCLAVABLE HAD AN ADAPTOR DISLODGED AND GOT STUCK IN THE LIGHT CABLE. THE REPORTED ISSUE WAS DISCOVERED AT THE USER FACILITY DURING MAINTENANCE OF THE SCOPE. THERE WAS NO PATIENT/USER HARM OR INJURY REPORTED DUE TO THE EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1584605 | TELESCOPE, 10 MM, 30°, HD, QUICK LOCK, AUTOCLAVABLE | RIGID OPTICAL LAPAROSCOPE | EWY | OLYMPUS WINTER & IBE GMBH¿ | WA53005A | 617246 | 04042761052764 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown |