MECTACER BIOLOX DELTA CERAMIC BALL HEAD 12/14 Ø 32 SIZE L +4
Report
- Report Number
- 3005180920-2023-00727
- Event Type
- Injury
- Date Received
- September 19, 2023
- Date of Event
- August 23, 2023
- Report Date
- September 19, 2023
- Manufacturer
- MEDACTA INTERNATIONAL SA
- Product Code
- LZO
- UDI-DI
- 07630030809170
- PMA / PMN Number
- K112115
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- US
- Reporter Occupation
- 003
Narratives
BATCH REVIEW PERFORMED ON 30 AUGUST 2023: LOT 2204275: 109 ITEMS MANUFACTURED AND RELEASED ON 19-MAY-2022. EXPIRATION DATE: 2027-05-05. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, 92 ITEMS OF THE SAME LOT HAVE BEEN SOLD WITHOUT ANY SIMILAR REPORTED EVENT DURING THE PERIOD OF REVIEW. ADDITIONAL DEVICE INVOLVED BATCH REVIEW PERFORMED ON 30 AUGUST 2023: LINER: MPACT 01.32.3244HCT FLAT PE HC LINER Ø32/E (K103721) LOT 2208320: 33 ITEMS MANUFACTURED AND RELEASED ON 07-JUNE-2022. EXPIRATION DATE: 2027-05-15. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, 29 ITEMS OF THE SAME LOT HAVE BEEN SOLD WITHOUT ANY SIMILAR REPORTED EVENT DURING THE PERIOD OF REVIEW.
AT ABOUT 7 MONTHS AFTER THE PRIMARY, THE PATIENT CAME IN DUE TO SIGNS OF AN INFECTION AND THE PATHOGEN IS UNKNOWN. THE SURGEON PERFORMED A WASHOUT AND REVISED THE HEAD AND LINER. THE SURGERY WAS COMPLETED SUCCESSFULLY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1891676 | MECTACER BIOLOX DELTA CERAMIC BALL HEAD 12/14 Ø 32 SIZE L +4 | HIP CERAMIC HEAD | LZO | MEDACTA INTERNATIONAL SA | 2204275 | 07630030809170 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 64 YR | Female | Required Intervention |