FDA Adverse Event Malfunction Summary report: N

BD BACTEC¿ LYTIC/10 ANAEROBIC/F CULTURE VIALS (PLASTIC)

MDR report key: 17769790 · Received September 18, 2023

Report

Report Number
3008352382-2023-00151
Event Type
Malfunction
Date Received
September 18, 2023
Date of Event
August 21, 2023
Report Date
October 14, 2023
Manufacturer
BECTON DICKINSON CARIBE LTD.
Product Code
MDB
UDI-DI
00382904420215
PMA / PMN Number
K123903
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
DE, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

H.6. INVESTIGATION SUMMARY: CATALOG: 442021. BATCH NO. 3130601. CUSTOMER REPORTED A MOLECULAR FALSE POSITIVE RESULT. NEITHER PHOTOS NOR RETURNED GOOD SAMPLES WERE RECEIVED. BD WAS UNABLE TO REPRODUCE THE CUSTOMER¿S EXPERIENCE WITH THE BACTEC PRODUCT BASED ON OUR INTERNAL PROCEDURES AND THE INTENDED USE OF THE PRODUCT. RETENTION SAMPLES WERE VISUALLY INSPECTED, TESTED FOR VIABLE CONTAMINATION BY SUB-CULTURING ON TSA, CHOCOLATE, SABOURAUD AND SCHAEDLER AGARS PLATES, VOLTAGE OUTPUT AND GRAM STAIN. ALL RESULTS WERE SATISFACTORY. BATCH HISTORY RECORD REVIEW DID NOT IDENTIFY ANY EVIDENCE FOR WHICH THE CUSTOMER SUBMITTED THE COMPLAINT. A COMPLAINT HISTORY REVIEW WAS CONDUCTED AND ONLY THE CURRENT COMPLAINT WAS FOUND RELATING TO THE INCIDENT LOT NUMBER AND THE ¿AS REPORTED¿ DEFECT CODE. COMPLAINT IS UNCONFIRMED BASED ON RETENTION SAMPLES AND BATCH HISTORY RECORD REVIEW RESULTS.

Additional Manufacturer Narrative · 0

H.3. A DEVICE EVALUATION AND/OR DEVICE HISTORY REVIEW IS ANTICIPATED BUT IS NOT COMPLETE. UPON COMPLETION, A SUPPLEMENTAL REPORT WILL BE FILED.

Description of Event or Problem · 0

REPORT 3 OF 3. IT WAS REPORTED THAT BD BACTEC¿ LYTIC/10 ANAEROBIC/F CULTURE VIALS (PLASTIC) MOLECULAR FALSE POSITIVES OCCURRED. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: BACTEC 442021 MOLECULAR FALSE POSITIVE WHAT RESULT DID THE CUSTOMER OBTAIN FROM THE BD PRODUCT? ONLY THE STAPH. EPI, E.COLI WHAT RESULT WAS THE CUSTOMER EXPECTING TO OBTAIN (IF DIFFERENT FROM THE OBTAINED RESULT) C. TROPICALIS ON SUBCULTURE. (B)(6): LOT? 3130601.

Description of Event or Problem · 0

REPORT 3 OF 3. IT WAS REPORTED THAT BD BACTEC¿ LYTIC/10 ANAEROBIC/F CULTURE VIALS (PLASTIC) MOLECULAR FALSE POSITIVES OCCURRED. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: BACTEC 442021 MOLECULAR FALSE POSITIVE. WHAT RESULT DID THE CUSTOMER OBTAIN FROM THE BD PRODUCT?: ONLY THE STAPH. EPI, ECOLI. WHAT RESULT WAS THE CUSTOMER EXPECTING TO OBTAIN (IF DIFFERENT FROM THE OBTAINED RESULT) C. TROPICALIS ON SUBCULTURE AUGUST 21: LOT? 3130601.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1948184 BD BACTEC¿ LYTIC/10 ANAEROBIC/F CULTURE VIALS (PLASTIC) SYSTEM, BLOOD CULTURING MDB BECTON DICKINSON CARIBE LTD. 3130601 00382904420215

Patients

Seq Age Sex Outcome Treatment
1 Unknown