AIMING ARM/ RADIOLUCENT
Report
- Report Number
- 8030965-2023-11728
- Event Type
- Malfunction
- Date Received
- September 18, 2023
- Date of Event
- January 1, 2023
- Manufacturer
- SYNTHES GMBH
- Product Code
- HWC
- UDI-DI
- 10886982297396
- PMA / PMN Number
- K201336
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
DEPUY SYNTHES IS SUBMITTING THIS REPORT PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH DEPUY SYNTHES HAS NOT BEEN ABLE TO INVESTIGATE OR VERIFY PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY FDA, DEPUY SYNTHES OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE DEVICE, DEPUY SYNTHES, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. IF THE INFORMATION IS UNKNOWN, NOT AVAILABLE OR DOES NOT APPLY, THE SECTION/FIELD OF THE FORM IS LEFT BLANK. D4: LOT AND EXPIRATION DATE. PART:03.043.029 LOT:2023566 MANUFACTURING SITE: WERK SELZACH SUPPLIER: (B)(4). RELEASE TO WAREHOUSE DATE: 25 NOV 2020 EXPIRATION DATE: NA. A MANUFACTURING RECORD EVALUATION WAS PERFORMED FOR THE FINISHED ARTICLE LOT AND NO NON-CONFORMANCES WERE IDENTIFIED. THE PRODUCT WAS RETURNED TO DEPUY SYNTHES FOR EVALUATION. THE DEPUY SYNTHES TEAM CONDUCTED A VISUAL INSPECTION OF THE RETURNED DEVICE. VISUAL ANALYSIS OF THE RETURNED DEVICE REVEALED THAT BOTH HOOKS OF THE LATCH OF AIM-ARM RADIOLUC, WERE BROKEN. THE BROKEN FRAGMENTS WERE NOT RETURNED. NO OTHER ISSUES WERE FOUND. DURING THE ANALYSIS OF THE COMPLAINTS TWO SUBCATEGORIES OF THE AIMING ARM (03.043.029) LATCH FAILURE WERE IDENTIFIED. THE FAILURE IS A BREAKAGE OF THE CARBON-FIBER REINFORCED LATCH. THE FAILURE MODE IS AN INTERLAMINAR BREAKAGE DUE TO SHEAR FORCES. BREAKAGE IS MOST LIKELY FAVORED BY DEFECTS IN THE STRUCTURE OF THE CARBON FIBER REINFORCED PEEK PLATE. IT IS IN THE NATURE OF THE MATERIAL THAT THE SHEAR STRENGTH IS HIGHLY ANISOTROPIC AND LOWEST BETWEEN CARBON FIBER LAYERS. LIKELY INTERLAMINAR SHEAR STRENGTH IS REDUCED BY DEFECTS RESULTING FORM MANUFACTURING ISSUES NOT LEADING TO COMPLETE BOND BETWEEN THE CARBON LAYERS. A DIMENSIONAL INSPECTION FOR THE AIM-ARM RADIOLUC WAS NOT PERFORMED DUE TO POST MANUFACTURING DAMAGE. AS PART OF DEPUY SYNTHES QUALITY PROCESS, ALL DEVICES ARE MANUFACTURED, INSPECTED, AND RELEASED TO APPROVED SPECIFICATIONS. THE OVERALL COMPLAINT WAS CONFIRMED AS THE OBSERVED CONDITION OF THE AIM-ARM RADIOLUC WOULD CONTRIBUTE TO THE COMPLAINED DEVICE ISSUE. ADDITIONAL MONITORING FOR ANY POTENTIAL SAFETY SIGNALS WILL BE CONDUCTED THROUGH COMPLAINT TRENDING AND OTHER POST-MARKET SAFETY SURVEILLANCE ACTIVITIES. DEVICE WAS USED FOR TREATMENT, NOT DIAGNOSIS. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE. G1: MANUFACTURING SITE NAME AND ADDRESS.
DEPUY SYNTHES IS SUBMITTING THIS REPORT PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH DEPUY SYNTHES HAS NOT BEEN ABLE TO INVESTIGATE OR VERIFY PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY FDA, DEPUY SYNTHES OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE DEVICE, DEPUY SYNTHES, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. IF THE INFORMATION IS UNKNOWN, NOT AVAILABLE OR DOES NOT APPLY, THE SECTION/FIELD OF THE FORM IS LEFT BLANK. H10 ADDITIONAL NARRATIVE: D2: ADDITIONAL PROCODE: JDS D9: COMPLAINANT PART IS EXPECTED TO BE RETURNED FOR MANUFACTURER REVIEW/INVESTIGATION, BUT HAS YET TO BE RECEIVED. E3: INITIAL REPORTER IS A SYNTHES EMPLOYEE. H3, H4, H6: WITHOUT A LOT NUMBER THE DEVICE HISTORY RECORDS REVIEW COULD NOT BE COMPLETED. PRODUCT WAS NOT RETURNED. BASED ON THE INFORMATION AVAILABLE, IT HAS BEEN DETERMINED THAT NO CORRECTIVE AND/OR PREVENTATIVE ACTION IS PROPOSED. THIS COMPLAINT WILL BE ACCOUNTED FOR AND MONITORED VIA POST MARKET SURVEILLANCE ACTIVITIES. IF ADDITIONAL INFORMATION IS MADE AVAILABLE, THE INVESTIGATION WILL BE UPDATED AS APPLICABLE. DEVICE WAS USED FOR TREATMENT, NOT DIAGNOSIS. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.
IT WAS REPORTED THAT ON AN UNKNOWN DATE IN 2023, TWO AIMING ARMS WERE DAMAGED DURING AN UNKNOWN PROCEDURE. THERE WAS NO REPORTED PATIENT IMPACT. NO FURTHER INFORMATION IS AVAILABLE. THIS REPORT IS FOR AN AIMING ARM/ RADIOLUCENT. THIS IS REPORT 2 OF 2 FOR (B)(4).
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 295246 | AIMING ARM/ RADIOLUCENT | SCREW, FIXATION, BONE | HWC | SYNTHES GMBH | 2023566 | 10886982297396 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown | AIMING ARM/ RADIOLUCENT |