FDA Adverse Event Malfunction Summary report: N

CARDIOSAVE HYBRID TYPE D PLUG

MDR report key: 17763924 · Received September 18, 2023

Report

Report Number
2249723-2023-04081
Event Type
Malfunction
Date Received
September 18, 2023
Report Date
April 3, 2024
Manufacturer
DATASCOPE CORP. - MAHWAH
Product Code
DSP
UDI-DI
10607567109008
PMA / PMN Number
K112372
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IN
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

A SUPPLEMENTAL REPORT WILL BE SUBMITTED UPON COMPLETION OF OUR INVESTIGATION.

Additional Manufacturer Narrative · 0

UPDATED FIELDS: B4,D9,E1(INITIAL REPORTER),E2,E3,G3,G6,H2,H3,H4,H6(TYPE OF INVESTIGATION, INVESTIGATION FINDINGS, COMPONENT CODES, INVESTIGATION CONCLUSIONS), H10, H11. CORRECTED FIELDS: D5. ADDITIONAL FIELDS: E1 (B)(6). IT WAS REPORTED THAT CARDIOSAVE INTRA-AORTIC BALLOON PUMP (IABP) CONTINUES BEEPING WHEN SWITCHING ON. A GETINGE FIELD SERVICE ENGINEER EVALUATED OBSERVED THAT SYSTEM IS SHOWING CONTINUOUS BEEP SOUND WHEN SWITCHING ON THE SYSTEM. PROBLEM SUSPECTED WITH POWER MANAGEMENT BOARD. REPLACED NEW POWER MANAGEMENT BOARD (D670-00-1162)AND UPDATED LATEST SOFTWARE. THEN CHECKED THE SYSTEM IT IS BOOTING UP BUT AFTER SOME TIME AUTOMATICALLY TRYING BOOT AGAIN AND AGAIN THEN FINALLY CONTINUES BEEPING. VIDEO GENERATOR BOARD KIT(D040-00-0456) REPLACED WHICH TOUCH SCREEN NOT WORKING PROPERLY. ALARM DATA BOARD(D670-00-0860) REPLACED THEN SYSTEM SHOWING ELECTRICAL TEST FAILURE #10 SO REPLACED FRONT END BOARD AND EXECUTIVE PROCESSOR BOARD (D670-00-0770) THEN PERFORMED ALL CALIBRATION. SWITCHING ON/OFF SEVERAL TIMES THEN CHECKED WITH DEMO BALLOON AND TRAINER IT IS WORKING FINE AND READY TO USE. IT IS UNKNOWN OF PATIENT INVOLVEMENT.

Description of Event or Problem · 0

IT WAS REPORTED THAT WHEN SWITCHING ON, THE CARDIOSAVE INTRA-AORTIC BALLOON PUMP (IABP) UNIT CONTINUES BEEPING.

Description of Event or Problem · 0

N/A.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
302784 CARDIOSAVE HYBRID TYPE D PLUG SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL DSP DATASCOPE CORP. - MAHWAH 0998-00-0800-33 10607567109008

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown