FDA Adverse Event Malfunction Summary report: N

DAVINCI XI

MDR report key: 17761040 · Received September 15, 2023

Report

Report Number
2955842-2023-18491
Event Type
Malfunction
Date Received
September 15, 2023
Date of Event
August 28, 2023
Report Date
August 28, 2023
Manufacturer
INTUITIVE SURGICAL, INC
Product Code
NAY
UDI-DI
00886874110720
PMA / PMN Number
K131861
Removal / Correction Number
N/A
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NH, US
Reporter Occupation
NURSE
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

AN INVESTIGATION WAS COMPLETED TO DETERMINE THE CAUSE OF THIS REPORTED EVENT. AN INTUITIVE SURGICAL, INC. (ISI) FIELD SERVICE ENGINEER (FSE) WAS DISPATCHED TO THE CUSTOMER SITE TO FURTHER INVESTIGATE THE REPORTED EVENT. BASED ON THE FIELD EVALUATION, THIS REPORTED EVENT WAS CONFIRMED. THE FSE FOUND A LOOSE SCREW ON THE RAIL, AND TIGHTENED IT TO RESOLVE THE ISSUE. THE SYSTEM WAS TESTED AND VERIFIED AS READY FOR USE. THE COMPLAINT WAS CONFIRMED BASED ON THE FIELD EVALUATION, WHICH INDICATES THAT THE DEVICE MAY HAVE CONTRIBUTED TO THE CUSTOMER REPORTED ISSUE.

Description of Event or Problem · 0

IT WAS REPORTED THAT DURING A DA VINCI-ASSISTED COMPONENT SEPARATION TAR SURGICAL PROCEDURE, UNIVERSAL SURGICAL MANIPULATOR (USM) 1 WAS STUCK AND COULD NOT BE ADVANCED. THE INTUITIVE SURGICAL INC. (ISI) TECHNICAL SUPPORT ENGINEER (TSE) ADVISED THE CUSTOMER TO DO AN EMERGENCY STOP AND RECOVER. THE CUSTOMER OPTED TO CONTINUE WITH USM 2, 3, AND 4. THE CUSTOMER LATER CALLED BACK AND STATED THAT THERE WAS A LOOSE SCREW ON USM 1. THERE WERE NO REPORTS OF PATIENT INJURY. THE PROCEDURE WAS COMPLETED ROBOTICALLY. INTUITIVE SURGICAL INC. (ISI) MADE MULTIPLE FOLLOW-UP ATTEMPTS TO OBTAIN ADDITIONAL INFORMATION. HOWEVER, NO FURTHER DETAILS HAVE BEEN RECEIVED AS OF THE DATE OF THIS REPORT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
299791 DAVINCI XI PATIENT SIDE CART, 4-ARM NAY INTUITIVE SURGICAL, INC 380652-46 N/A 00886874110720

Patients

Seq Age Sex Outcome Treatment
1 Unknown DA VINCI INSTRUMENTS AND ACCESSORIES