FDA Adverse Event Death Summary report: N

MITRACLIP G4 CLIP DELIVERY SYSTEM (MDD)

MDR report key: 17760848 · Received September 15, 2023

Report

Report Number
2135147-2023-04012
Event Type
Death
Date Received
September 15, 2023
Date of Event
August 23, 2023
Report Date
November 3, 2023
Manufacturer
ABBOTT MEDICAL
Product Code
NKM
PMA / PMN Number
P100009
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

INVESTIGATION IS NOT YET COMPLETE. A FOLLOW-UP REPORT WILL BE SUBMITTED WITH ALL ADDITIONAL RELEVANT INFORMATION.

Additional Manufacturer Narrative · 0

THE DEVICE WAS NOT RETURNED FOR ANALYSIS. A REVIEW OF THE LOT HISTORY RECORD REVEALED NO MANUFACTURING NONCONFORMITIES ISSUED TO THE REPORTED LOT THAT WOULD HAVE CONTRIBUTED TO THIS EVENT. ADDITIONALLY, A REVIEW OF THE COMPLAINT HISTORY IDENTIFIED NO SIMILAR COMPLAINTS REPORTED FROM THIS LOT. BASED ON AVAILABLE INFORMATION, THE REPORTED HYPOTENSION (THERAPY/NON-SURGICAL TREATMENT, ADDITIONAL) APPEARS TO BE DUE TO THE CARDIAC TAMPONADE. THE REPORTED CARDIAC TAMPONADE (TREATMENT WITH MEDICATION(S)), AND THE REPORTED HEMORRHAGE/BLOOD LOSS/BLEEDING (BLOOD TRANSFUSION) ASSOCIATED WITH THE BLOOD LOSS, WERE DUE TO THE PERICARDIAL EFFUSION. THE CAUSE OF THE PERICARDIAL EFFUSION (THERAPY/NON-SURGICAL TREATMENT, ADDITIONAL) COULD NOT BE DETERMINED. THE CAUSE OF THE REPORTED UNINTENDED MOVEMENT (SLEEVE STEERING ISSUES) ASSOCIATED WITH THE DC MOVING IN THE OPPOSITE DIRECTION THAN EXPECTED COULD NOT BE DETERMINED. THE CAUSE OF THE REPORTED DEATH / EXPIRED (CARDIAC DEATH) COULD NOT BE DETERMINED. THE REPORTED IMAGE RESOLUTION POOR WAS ASSOCIATED WITH VISUALIZATION DIFFICULTY. THE REPORTED PATIENT EFFECTS OF DEATH, CARDIAC TAMPONADE, HYPOTENSION, HEMORRHAGE, AND PERICARDIAL EFFUSION, AS LISTED IN THE MITRACLIP SYSTEM INSTRUCTIONS FOR USE, ARE KNOWN POSSIBLE COMPLICATIONS ASSOCIATED WITH MITRACLIP PROCEDURES. THE REPORTED UNEXPECTED MEDICAL INTERVENTION AND MEDICATION REQUIRED WERE RESULTS OF CASE-SPECIFIC CIRCUMSTANCES. THERE IS NO INDICATION OF A PRODUCT QUALITY ISSUE WITH RESPECT TO MANUFACTURE, DESIGN, OR LABELING.

Description of Event or Problem · 0

THIS IS FILED TO REPORT A DEATH. IT WAS REPORTED THAT A MITRACLIP PROCEDURE WAS PERFORMED TO TREAT MIXED MITRAL REGURGITATION (MR) GRADE 3-4. TRANSESOPHAGEAL ECHOCARDIOGRAM (TEE) IMAGING WAS OF POOR QUALITY. HEPARIN WAS ADMINISTERED IMMEDIATELY AFTER THE TRANSSEPTAL PUNCTURE, MAINTAINING AN ACTIVATED CLOTTING TIME (ACT) LEVEL 250 SEC. A XTW CLIP WAS INSERTED, BUT WHEN IN THE LEFT ATRIUM IT APPEARED TO BE MOVING IN OPPOSITE DIRECTIONS THAN THE APPLIED KNOB ON FLUOROSCOPY. IT WAS THOUGHT THE CLIP MAY HAVE BEEN MISALIGNED IN THE STEERABLE GUIDE CATHETER. WHEN INSERTING THE CDS INTO THE SGC, THE PHYSICIAN REPORTED THAT THE LONGITUDINAL MARKER ON THE STEERABLE SLEEVE SHAFT SHAFT ALIGNED WITH THE ALIGNMENT MARKER ON THE GUIDE HEMOSTASIS VALVE. THE CLIP WAS PULLED BACK TO THE SGC TO RE-ALIGN, BUT THE PATIENT SUFFERED HEMODYNAMIC COLLAPSE - DIRECT BLOOD PRESSURE DROP TO SYSTOLIC 40MMHG, AND A MASSIVE PERICARDIAL EFFUSION WAS IMMEDIATELY NOTED. DUE TO THE POOR IMAGING, IT WAS UNKNOWN IF THE CLIP CONTACTED ANATOMY AND IT WAS UNKNOWN WHERE THE EXACT LOCATION OF THE EFFUSION WAS. THE CLIP WAS REMOVED AND THE SGC WAS MOVED TO THE RIGHT ATRIUM. PERICARDIOCENTESIS WAS PERFORMED ALONG WITH ADMINISTERING OF PROTAMIN 7.500IE, AND BLOOD TRANSFUSIONS. THE PERICARDIAL EFFUSION DEVELOPED INTO CARDIAC TAMPONADE. PATIENT RECEIVED MEDICAL RESUSCITATION WITH ALL EMERGENCY MEDICATIONS - SUPRARENIN 1:10 ML AT THE BEGINNING TO SUPRARENIN PERFUSOR PURE. DESPITE THE EFFORTS, THE PATIENT EVENTUALLY DIED UNDER RESUSCITATIVE MEASURES 35 MINUTES AFTER THE ONSET OF THE PERICARDIAL EFFUSION. IT CANT BE RULED OUT THAT THE MITRACLIP SYSTEM MAY HAVE CAUSED TISSUE DAMAGE OR CONTRIBUTED TO THE PERICARDIAL EFFUSION DUE TO POOR IMAGING. NO ADDITIONAL INFORMATION WAS PROVIDED.

Description of Event or Problem · 0

IT WAS REPORTED THAT A MITRACLIP PROCEDURE WAS PERFORMED TO TREAT MIXED MITRAL REGURGITATION (MR) GRADE 3-4. TRANSESOPHAGEAL ECHOCARDIOGRAM (TEE) IMAGING WAS OF POOR QUALITY. HEPARIN WAS ADMINISTERED IMMEDIATELY AFTER THE TRANSSEPTAL PUNCTURE, MAINTAINING AN ACTIVATED CLOTTING TIME (ACT) LEVEL 250 SEC. A XTW CLIP WAS INSERTED, BUT WHEN IN THE LEFT ATRIUM IT APPEARED TO BE MOVING IN OPPOSITE DIRECTIONS THAN THE APPLIED KNOB ON FLUOROSCOPY. WHEN THE M-KNOB WAS APPLIED THE CLIP DELIVERY SYSTEM (CDS) MOVED TOWARDS THE ATRIAL ROOF. IT WAS UNKNOWN IF THE CDS WAS MISALIGNED IN THE STEERABLE GUIDE CATHETER (SGC). WHEN INSERTING THE CDS INTO THE SGC, THE PHYSICIAN REPORTED THAT THE LONGITUDINAL MARKER ON THE STEERABLE SLEEVE SHAFT ALIGNED WITH THE ALIGNMENT MARKER ON THE GUIDE HEMOSTASIS VALVE. THE CLIP WAS PULLED BACK TO THE SGC TO RE-ALIGN, BUT THE PATIENT SUFFERED HEMODYNAMIC COLLAPSE - DIRECT BLOOD PRESSURE DROP TO SYSTOLIC 40MMHG, AND A MASSIVE PERICARDIAL EFFUSION WAS IMMEDIATELY NOTED. DUE TO THE POOR IMAGING, IT WAS UNKNOWN IF THE CLIP CONTACTED ANATOMY AND IT WAS UNKNOWN WHERE THE EXACT LOCATION OF THE EFFUSION WAS. THE CLIP WAS REMOVED AND THE SGC WAS MOVED TO THE RIGHT ATRIUM. PERICARDIOCENTESIS WAS PERFORMED ALONG WITH ADMINISTERING OF PROTAMIN 7.500IE, AND BLOOD TRANSFUSIONS. THE PERICARDIAL EFFUSION DEVELOPED INTO CARDIAC TAMPONADE. PATIENT RECEIVED MEDICAL RESUSCITATION WITH ALL EMERGENCY MEDICATIONS - SUPRARENIN 1:10 ML AT THE BEGINNING TO SUPRARENIN PERFUSOR PURE. DESPITE THE EFFORTS, THE PATIENT EVENTUALLY DIED UNDER RESUSCITATIVE MEASURES 35 MINUTES AFTER THE ONSET OF THE PERICARDIAL EFFUSION. IT CANT BE RULED OUT THAT THE MITRACLIP SYSTEM MAY HAVE CAUSED TISSUE DAMAGE OR CONTRIBUTED TO THE PERICARDIAL EFFUSION DUE TO POOR IMAGING. NO ADDITIONAL INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1040091 MITRACLIP G4 CLIP DELIVERY SYSTEM (MDD) MITRAL VALVE REPAIR DEVICES NKM ABBOTT MEDICAL 30606R3054

Patients

Seq Age Sex Outcome Treatment
1 Unknown Death STEERABLE GUIDE CATHETER