VITROS CHEMISTRY PRODUCTS NA+ SLIDES
Report
- Report Number
- 0001319809-2023-00105
- Event Type
- Malfunction
- Date Received
- September 15, 2023
- Date of Event
- August 21, 2023
- Report Date
- September 15, 2023
- Manufacturer
- ORTHO-CLINICAL DIAGNOSTICS, INC.
- Product Code
- JGS
- UDI-DI
- 10758750004812
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AL, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
THE INVESTIGATION DETERMINED THAT A HIGHER THAN EXPECTED SODIUM RESULT WAS OBTAINED FROM A COLLEGE OF AMERICAN PATHOLOGISTS (CAP). PROFICIENCY SAMPLE REPEAT PROCESSED USING VITROS CHEMISTRY PRODUCTS SODIUM (NA+) SLIDES, LOT 4211-1096-5681 ON A VITROS XT 7600 INTEGRATED SYSTEM. THE ASSIGNABLE CAUSE OF THE EVENT CANNOT BE DETERMINED. AN ISSUE WITH THE AGE OF THE CAP SAMPLE CANNOT BE RULED OUT AS A CONTRIBUTOR TO THE EVENT. PER CAP, OPENED SPECIMENS ARE STABLE AT 2-8 DEGREES CELSIUS FOR 7 DAYS. THE CUSTOMER INITIALLY PROCESSED THE SAMPLE IN JUNE 2023 AND REPEATED THE SAMPLE IN AUGUST 2023, OBTAINING THE HIGHER THAN EXPECTED RESULT ABOVE. HISTORICAL QUALITY CONTROL RESULTS FOR VITROS NA+ LOT 4211-1096-5681 SHOWED SOME IMPRECISION AT HIGHER CONCENTRATIONS, HOWEVER CURRENT PERFORMANCE WAS WITHIN EXPECTATION, INDICATING THAT A VITROS NA+ LOT 4211-1096-5681 PERFORMANCE ISSUE IS NOT A LIKELY CONTRIBUTOR TO THE EVENT. CONTINUAL TRACKING AND TRENDING OF COMPLAINTS HAS NOT IDENTIFIED ANY SIGNALS THAT WOULD POINT TO A POTENTIAL SYSTEMIC ISSUE WITH VITROS NA+ LOT 4211-1096-5681. AS PRECISION TESTING WAS NOT RUN AT THE TIME OF THE EVENT, AN INSTRUMENT RELATED ISSUE COULD NOT BE CONFIRMED NOR RULED OUT AS A CONTRIBUTING FACTOR OF THE EVENTS.
A CUSTOMER CONTACTED THE ORTHO CLINICAL DIAGNOSTICS (ORTHO) TECHNICAL SOLUTION CENTER (TSC) TO REPORT A HIGHER THAN EXPECTED SODIUM RESULT WAS OBTAINED FROM A COLLEGE OF AMERICAN PATHOLOGISTS (CAP) PROFICIENCY SAMPLE REPEAT PROCESSED USING VITROS CHEMISTRY PRODUCTS SODIUM (NA+) SLIDES LOT 4211-1096-5681 ON A VITROS XT 7600 INTEGRATED SYSTEM. CAP CHM-07 NA+ RESULT OF 135.8 MMOL/L VERSUS THE EXPECTED RESULT OF 129.8 BIASED RESULTS OF THE MAGNITUDE AND DIRECTION OBSERVED MAY LEAD TO INAPPROPRIATE PHYSICIAN ACTION IF THEY WERE TO OCCUR UNDETECTED ON PATIENT SAMPLES. THE UNEXPECTED RESULT WAS FROM A PROFICIENCY SAMPLE REPEAT. THERE WAS NO ALLEGATION OF PATIENT HARM AS A RESULT OF THIS EVENT. THIS REPORT CORRESPONDS TO ORTHO CLINICAL DIAGNOSTICS INC (ORTHO) COMPLAINT NUMBER: (B)(4) AND REPORTABILITY ASSESSMENT: 603169.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 328638 | VITROS CHEMISTRY PRODUCTS NA+ SLIDES | IN-VITRO DIAGNOSTICS | JGS | ORTHO-CLINICAL DIAGNOSTICS, INC. | 8379034 | 4211-1096-5681 | 10758750004812 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown |