FDA Adverse Event Malfunction Summary report: N

HARMONIC ACE PLUS 7 ADV HEMOSTASIS 36

MDR report key: 17756418 · Received September 15, 2023

Report

Report Number
3005075853-2023-06665
Event Type
Malfunction
Date Received
September 15, 2023
Date of Event
June 21, 2023
Report Date
September 15, 2023
Manufacturer
ETHICON ENDO-SURGERY, LLC.
Product Code
LFL
UDI-DI
10705036014621
PMA / PMN Number
K132612
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FR
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 0

(B)(4). DATE SENT: 9/15/2023. D4 BATCH #: W7036C. INVESTIGATION SUMMARY: THE PRODUCT WAS RETURNED FOR EVALUATION. VISUAL INSPECTION AND FUNCTIONAL TESTING WERE CONDUCTED ON THE RETURNED DEVICE. VISUAL ANALYSIS OF THE RETURNED SAMPLE DETERMINED THAT THE DEVICE WAS RETURNED WITH THE DISTAL TIP OF THE BLADE BROKEN OFF AND NOT RETURNED. THE REMAINING BLADE PORTION WAS SCRATCHED, AND WITH EVIDENCE OF CONTACT WITH METAL IN OR OUT OF THE OPERATIVE FIELD. DURING FUNCTIONAL TESTING ON GEN11, AN ALERT SCREEN WAS DISPLAYED. A PROBABLE CAUSE FOR THE DEVICE TO STOP ACTIVATING AND THE GEN11 TO DISPLAY AN ALERT SCREEN IS BLADE DAMAGE. THE DEVICE WAS DISASSEMBLED TO INSPECT THE INTERNAL COMPONENTS AND NO ANOMALIES WERE FOUND.  PROBABLE CAUSES OF BLADE DAMAGE, INCLUDING BREAKAGE, ARE EXTERNAL CONTACT DURING PRE-OP OR GENERAL USE, BLADE CONTACT WITH OTHER DEVICES, STAPLES OR CLIPS DURING THE PROCEDURE OR USING ANY MEANS OTHER THAN THE BLADE WRENCH TO ATTACH OR DETACH THE BLADE.   ONCE MINOR BLADE DAMAGE HAS OCCURRED, SUBSEQUENT ACTIVATIONS MAY INCREASE THE SEVERITY OF THE BLADE DAMAGE. THIS IN TURN CAN RESULT IN ACTIVATION ISSUES SUCH AS FAILING THE PRE-RUN TEST WITH THE GENERATOR AND DISPLAYING AN ALERT SCREEN. THESE ALERT SCREENS THAT CAN RESULT ARE SUCH AS ¿TIGHTEN ASSEMBLY¿, ¿BLADE ERROR DETECTED¿ OR "RELAXED PRESSURE ON BLADE" FOLLOWED BY A ¿REPLACE INSTRUMENT¿ SCREEN LATER IN THE PROCEDURE. CONTINUED USAGE OF THE DAMAGE BLADE CAN RESULT IN A BROKEN BLADE. AS PART OF THE QUALITY PROCESS ALL DEVICES ARE MANUFACTURED, INSPECTED, AND RELEASED TO APPROVED SPECIFICATIONS.  A MANUFACTURING RECORD EVALUATION WAS PERFORMED FOR THE FINISHED DEVICE BATCH W7036C, AND NO NON-CONFORMANCES WERE IDENTIFIED.

Description of Event or Problem · 0

IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE THERE WAS A FAULTY DEVICE DURING THE INTERVENTION. CHANGE OF DEVICE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
779772 HARMONIC ACE PLUS 7 ADV HEMOSTASIS 36 INSTRUMENT, ULTRASONIC SURGICAL LFL ETHICON ENDO-SURGERY, LLC. 10705036014621

Patients

Seq Age Sex Outcome Treatment
1 Unknown HANDPIECE AND GENERATOR