FDA Adverse Event Malfunction Summary report: N

UNKNOWN

MDR report key: 17754149 · Received September 15, 2023

Report

Report Number
3002808486-2023-00236
Event Type
Malfunction
Date Received
September 15, 2023
Report Date
December 14, 2023
Manufacturer
WILLIAM COOK EUROPE
Product Code
DTK
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SC, US
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

MANUFACTURER REF# (B)(4). INVESTIGATION IS STILL IN PROGRESS. THIS REPORT IS REQUIRED BY THE FDA UNDER 21 CFR PART 803. THIS REPORT IS BASED ON UNCONFIRMED INFORMATION SUBMITTED BY OTHERS. NEITHER THE SUBMISSION OF THIS REPORT NOR ANY STATEMENT MADE IN IT IS INTENDED TO BE AN ADMISSION THAT ANY COOK DEVICE IS DEFECTIVE OR MALFUNCTIONED; THAT A DEATH OR SERIOUS INJURY OCCURRED; OR THAT ANY COOK DEVICE CAUSED OR CONTRIBUTED TO; OR IS LIKELY TO CAUSE OR CONTRIBUTE TO A DEATH OR SERIOUS INJURY IF A MALFUNCTION OCCURRED.

Additional Manufacturer Narrative · 0

MANUFACTURER¿S REF# (B)(4). D1: DEVICE REFERED TO AS A GUNTHER TULIP FILTER. G4: DEVICE IS APPROVED UNDER 510 (K) K211874. INVESTIGATION IS STILL IN PROGRESS: THIS REPORT IS REQUIRED BY THE FDA UNDER 21 CFR PART 803. THIS REPORT IS BASED ON UNCONFIRMED INFORMATION SUBMITTED BY OTHERS. NEITHER THE SUBMISSION OF THIS REPORT NOR ANY STATEMENT MADE IN IT IS INTENDED TO BE AN ADMISSION THAT ANY COOK DEVICE IS DEFECTIVE OR MALFUNCTIONED; THAT A DEATH OR SERIOUS INJURY OCCURRED; OR THAT ANY COOK DEVICE CAUSED OR CONTRIBUTED TO; OR IS LIKELY TO CAUSE OR CONTRIBUTE TO A DEATH OR SERIOUS INJURY IF A MALFUNCTION OCCURRED.

Additional Manufacturer Narrative · 0

MANUFACTURER REF# (B)(4). SUMMARY OF INVESTIGATIONAL FINDINGS: THE TULIP FILTER HAD PENETRATED TWO YEARS AFTER PLACEMENT. THE FILTER WAS RETRIEVED, BUT NO OTHER INFORMATION COULD BE OBTAINED. THE FILTER WAS NOT RETURNED AND NO IMAGING COULD BE PROVIDED. THEREFORE, BASED ON THE LIMITED INFORMATION ONLY IT WOULD BE INAPPROPRIATE TO SPECULATE AT WHAT MAY OR MAY NOT HAVE CAUSED THE FILTER TO PENETRATE THE IVC TWO YEARS AFTER PLACEMENT. FILTER INTERACTS WITH IVC WALL, E.G. PENETRATION/PERFORATION/EMBEDMENT. THIS MAY BE EITHER SYMPTOMATIC OR ASYMPTOMATIC. POTENTIAL CAUSES MAY INCLUDE IMPROPER DEPLOYMENT; AND (OR) EXCESSIVE FORCE OR MANIPULATIONS NEAR AN IN-SITU FILTER (E.G., A SURGICAL OR ENDOVASCULAR PROCEDURE IN THE VICINITY OF A FILTER). POTENTIAL ADVERSE EVENTS THAT MAY OCCUR INCLUDE, BUT ARE NOT LIMITED TO, THE FOLLOWING: VASCULAR TRAUMA, VENA CAVA PERFORATION, VENA CAVA PENETRATION. THERE ARE ADEQUATE CONTROLS IN PLACE TO ENSURE THAT THIS TYPE OF DEVICE IS MANUFACTURED TO SPECIFICATIONS. COOK WAS UNABLE TO CONDUCT A REVIEW OF THE DEVICE HISTORY RECORD, AS THE LOT NUMBER OF THE COMPLAINT DEVICE WAS NOT PROVIDED FOR THE INVESTIGATION. COOK MEDICAL WILL CONTINUE TO MONITOR FOR SIMILAR EVENTS. THIS REPORT IS REQUIRED BY THE FDA UNDER 21 CFR PART 803. THIS REPORT IS BASED ON UNCONFIRMED INFORMATION SUBMITTED BY OTHERS. NEITHER THE SUBMISSION OF THIS REPORT NOR ANY STATEMENT MADE IN IT IS INTENDED TO BE AN ADMISSION THAT ANY COOK DEVICE IS DEFECTIVE OR MALFUNCTIONED; THAT A DEATH OR SERIOUS INJURY OCCURRED; OR THAT ANY COOK DEVICE CAUSED OR CONTRIBUTED TO; OR IS LIKELY TO CAUSE OR CONTRIBUTE TO A DEATH OR SERIOUS INJURY IF A MALFUNCTION OCCURRED.

Description of Event or Problem · 0

ADDITIONAL INFORMATION RECEIVED 18OCT2023: THE FILTER WAS RETRIEVED.

Description of Event or Problem · 0

DESCRIPTION OF EVENT ACCORDING TO INITIAL REPORTER: THE IMPLANTING PHYSICIAN MENTIONED TO HIS COLLEAGUE THAT A PATIENT THAT HAD AN IVC FILTER IMPLANTED AROUND 2 YEARS AGO AT AN UNKNOWN FACILITY CAME TO THEIR FACILITY AND EXHIBITED A PENETRATION. THE OTHER PHYSICIAN REPORTED THIS SECOND-HAND INFORMATION TO THE COOK REP. THE PHYSICIAN SAID THAT THE FILTER WAS PLACED DUE TO TRAUMA. NO FUTURE INTERVENTIONS HAVE BEEN CONFIRMED.

Description of Event or Problem · 0

NO ADDITIONAL INFORMATION REGARDING THE PATIENT AND/OR EVENT HAS BEEN RECEIVED SINCE THE PREVIOUS MEDWATCH REPORT WAS SENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
294503 UNKNOWN DTK FILTER, INTRAVASCULAR, CARDIOVASCULAR DTK WILLIAM COOK EUROPE

Patients

Seq Age Sex Outcome Treatment
1 Unknown