FDA Adverse Event Malfunction Summary report: N

PENTAX

MDR report key: 17753131 · Received September 15, 2023

Report

Report Number
9610877-2023-58917
Event Type
Malfunction
Date Received
September 15, 2023
Date of Event
September 7, 2023
Report Date
September 15, 2023
Manufacturer
HOYA CORPORATION PENTAX TOKYO OFFICE
Product Code
EOB
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
UK
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

THIS DEVICE IS CLASSIFIED AS IMPORT FOR EXPORT, THEREFORE 510K IS NOT APPLICABLE. MODEL FNL-10RP3 IS AVAILABLE IN THE USA WITH A 510K NUMBER K951196. THE PRODUCT WAS RETURNED TO PENTAX MEDICAL FOR REPAIR. OUR TECHNICIAN CHECKED THE RETURNED UNIT AND CONFIRMED THAT THE OCULAR FLUID DAMAGE. BASED ON THE RESULT, WE CONCLUDED THAT IT WAS CAUSED DUE TO THE FLUID DAMAGE FROM THE OCULAR. BASED ON THE TECHNICAL REPORT "HR-RPT-0586(IMAGE FAILURE)" AND/OR THE RISK ANALYSIS RESULTS, IT WAS EVALUATED TO SUBMIT MDR.

Description of Event or Problem · 0

THE TIME OF EVENT IS NOT DURING PROCEDURE. THERE WAS NO REPORT OF PATIENT HARM. FIBER IMAGE FAILURE (FLUID DAMAGE).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
686398 PENTAX FIBER NASO PHARYNGO LARYNGOSCOPE EOB HOYA CORPORATION PENTAX TOKYO OFFICE FNL-10RBS

Patients

Seq Age Sex Outcome Treatment
1 Unknown