FDA Adverse Event Malfunction Summary report: N

STANDARD WISHBONE ASSEMBLY

MDR report key: 17749812 · Received September 14, 2023

Report

Report Number
3004608878-2023-00170
Event Type
Malfunction
Date Received
September 14, 2023
Date of Event
August 23, 2023
Report Date
October 4, 2023
Manufacturer
INTEGRA LIFESCIENCES CORPORATION OH/USA
Product Code
GAD
PMA / PMN Number
EXEMPT
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER
Health Professional
N

Narratives

Additional Manufacturer Narrative · 0

THE STANDARD WISHBONE ASSEMBLY (4020) WAS RETURNED FOR EVALUATION: FAILURE ANALYSIS: UPON INSPECTION OF THE RETURNED UNIT, IT WAS DISCOVERED THAT THE ARMS INDEED MOVED IN THE LOCKED POSITION. PREVENTATIVE MAINTENANCE AND REPLACEMENT OF WORN PARTS WERE REQUIRED. ROOT CAUSE ANALYSIS: COMPLAINT CONFIRMED. BASED ON THE EVALUATION BY INTEGRA SERVICE AND REPAIR, THE ROOT CAUSE WAS MOST LIKELY DUE TO INSUFFICIENT ROUTINE PREVENTIVE MAINTENANCE. BASED ON THIS EVALUATION, ROOT CAUSE DETERMINATION, AND RISK ASSESSMENT, NO FURTHER ACTION OR INVESTIGATION IS PLANNED.

Additional Manufacturer Narrative · 0

AN INVESTIGATION HAS BEEN INITIATED BASED ON THE REPORTED INFORMATION. UPON COMPLETION OF THE INVESTIGATION, A FOLLOW-UP REPORT WILL BE SUBMITTED.

Description of Event or Problem · 0

N/A.

Description of Event or Problem · 0

IT WAS REPORTED THAT THE ARM OF STANDARD WISHBONE ASSEMBLY (4020) WAS NOT HOLDING PROPERLY; IT WAS ROTATING. ONCE THE WISHBONE FRAME WAS LOCKED, IT IS MEANT TO STAY LOCKED; HOWEVER, THE WISHBONE FRAME MOVED. A 30-MINUTE DELAY IN SURGERY OCCURRED AS THE SURGEON HAD TO WAIT FOR ANOTHER RETRACTOR. A NEW RETRACTOR FRAME WAS OPENED, AND THE SURGEON TOLD STAFF THAT THE REPLACEMENT RETRACTOR WAS ALSO MOVING; HOWEVER, THE SURGEON USED THAT FRAME BECAUSE THERE WAS NO REPLACEMENT AVAILABLE. IT WAS INDICATED THAT THE REPLACEMENT RETRACTOR DID NOT CONTRIBUTE TO A DELAY IN SURGERY. NO ADVERSE EVENT WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
362427 STANDARD WISHBONE ASSEMBLY SURGICAL RETRACTOR GAD INTEGRA LIFESCIENCES CORPORATION OH/USA M42694

Patients

Seq Age Sex Outcome Treatment
1 Unknown