FDA Adverse Event Malfunction Summary report: N

ARCHITECT I2000SR

MDR report key: 17748211 · Received September 14, 2023

Report

Report Number
3016438761-2023-00498
Event Type
Malfunction
Date Received
September 14, 2023
Date of Event
September 4, 2023
Report Date
September 14, 2023
Manufacturer
ABBOTT LABORATORIES
Product Code
JJE
UDI-DI
00380740006235
PMA / PMN Number
EXEMPT
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CH
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

DURING THE REQUESTED SITE VISIT, THE ABBOTT FIELD SERVICE ENGINEER (FSE) CLEANED THE SHUTTER (ROHS)_PART NUMBER 7-78200-02 WHICH RESOLVED THE ISSUE. A REVIEW OF THE (B)(6) SERVICE HISTORY WAS PERFORMED AND NO ADDITIONAL ERRATIC RESULTS PRE- OR POST THE CURRENT COMPLAINT WERE IDENTIFIED. THE ARCHITECT SYSTEM OPERATIONS MANUAL ADDRESSES OPERATIONAL PRECAUTIONS AND LIMITATIONS AND TROUBLESHOOTING OF THE FLUIDICS SUBSYSTEM. A REVIEW OF HISTORICAL DATA REVEALED NO TRENDS, SYSTEMIC ISSUES, OR RELATED NONCONFORMANCES. THE ARCHITECT SYSTEM OPERATIONS MANUAL ADDRESSES OPERATIONAL PRECAUTIONS AND LIMITATIONS AND TROUBLESHOOTING OF THE FLUIDICS SUBSYSTEM AND COMPONENT REPLACEMENT, REMOVAL AND VERIFICATION OF THE PARTS. A REVIEW OF TRACKING AND TRENDING OF THE ARCHITECT I2000SR PROCESSING MODULE DID NOT IDENTIFY ANY TRENDS ASSOCIATED WITH THE COMPLAINT ISSUE. A REVIEW OF TRACKING AND TRENDING DID NOT IDENTIFY ANY TRENDS FOR THE SHUTTER. A REVIEW OF THE PRODUCT LABELING CONCLUDED THAT THE ISSUE IS SUFFICIENTLY ADDRESSED. BASED ON THE INVESTIGATION NO PRODUCT DEFICIENCY WAS IDENTIFIED FOR EITHER THE ARCHITECT I2000SR, SN (B)(6), OR THE SHUTTER.

Description of Event or Problem · 0

THE CUSTOMER OBSERVED FALSELY ELEVATED ARCHITECT TROPONIN RESULT ON ONE FEMALE PATIENT. THE RESULT PROVIDED WERE: FIRST RESULT WITH SERIAL NUMBER (B)(6) /SECOND RESULTS WITH SERIAL NUMBER (B)(6) FROM SAME LOT NUMBER OF TROPONIN REAGENT. PATIENT 5: = 24.4 PG/ML /=2.0 PG/ML. LABORATORY REFERENCE RANGE FOR TROPONIN; FEMALE=0.0 TO 15.6 PG/ML; MEN=0.0 TO 34.2 PG/ML THERE WAS NO REPORTED IMPACT TO PATIENT MANAGEMENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2282684 ARCHITECT I2000SR ANALYZER, CHEMISTRY (PHOTOMETRIC, DISCRETE), FOR CLINICAL USE JJE ABBOTT LABORATORIES 00380740006235

Patients

Seq Age Sex Outcome Treatment
1 Female ARC I2000SR INST, 03M74-02, (B)(6)