IDYS-TLIF TIVAC
Report
- Report Number
- 3009962553-2023-00003
- Event Type
- Malfunction
- Date Received
- September 14, 2023
- Date of Event
- August 15, 2023
- Report Date
- September 13, 2023
- Manufacturer
- CLARIANCE SAS
- Product Code
- OVD
- UDI-DI
- 03700780628643
- PMA / PMN Number
- K183259
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MO, US
- Reporter Occupation
- 003
Narratives
WHILE IMPACTING THE TLIF CAGE INTO THE DISC SPACE, BEFORE ARTICULATING THE INSERTER, THE PROXIMAL TANTALUM MARKER PORTION OF THE CAGE BROKE FROM THE REST OF THE CAGE. THE SURGEON WAS ABLE TO RETRIEVE THE BROKEN PORTION AND REMOVE IT FROM THE DISC SPACE. THEN, THE SURGEON WAS ABLE TO POSITION THE REST OF THE CAGE ACCORDINGLY. THIS EVENT IS REPORTABLE AS A MALFUNCTION.
WHILE IMPACTING THE TLIF CAGE INTO THE DISC SPACE, BEFORE ARTICULATING THE INSERTER, THE PROXIMAL TANTALUM MARKER PORTION OF THE CAGE BROKE FROM THE REST OF THE CAGE. THE SURGEON WAS ABLE TO RETRIEVE THE BROKEN PORTION AND REMOVE IT FROM THE DISC SPACE. THEN, THE SURGEON WAS ABLE TO POSITION THE REST OF THE CAGE ACCORDINGLY. THIS EVENT IS REPORTABLE AS A MALFUNCTION. ON 11/28/2023, THIS MEDWATCH 3500A FILING NOW INCLUDES THE ANALYSIS REPORT FROM THE MANUFACTURER.
THE SURGEON WAS ON THE PATIENT'S RIGHT AT L4-5 USING A 9MM TI COATED PEEK BANANA CAGE WITH THE 25-DEGREE ANGLED INSERTER. WHILE IMPACTING INTO THE DISC SPACE, BEFORE ARTICULATING THE INSERTER, THE PROXIMAL TANTALUM MARKER PORTION OF THE CAGE BROKE FROM THE REST OF THE CAGE. HE WAS ABLE TO RETRIEVE THE BROKEN PORTION AND REMOVE IT FROM THE DISC SPACE. HE THEN USED A TAMP/PUSHER TO POSITION THE REST OF THE CAGE ACCORDINGLY. THERE WAS NOT A SIGNIFICANT DELAY TO THE CASE, OR ANY HARM DONE TO THE PATIENT BECAUSE OF THE INCIDENT. ALL THE BROKEN PIECES OF THE IMPLANT WERE REMOVED FROM THE PATIENT AND DISCARDED BY THE SURGICAL STAFF AND CAN'T BE SENT BACK FOR ANALYSIS.
THE SURGEON WAS ON THE PATIENT'S RIGHT AT L4-5 USING A 9MM TI COATED PEEK BANANA CAGE WITH THE 25-DEGREE ANGLED INSERTER. WHILE IMPACTING INTO THE DISC SPACE, BEFORE ARTICULATING THE INSERTER, THE PROXIMAL TANTALUM MARKER PORTION OF THE CAGE BROKE FROM THE REST OF THE CAGE. HE WAS ABLE TO RETRIEVE THE BROKEN PORTION AND REMOVE IT FROM THE DISC SPACE. HE THEN USED A TAMP/PUSHER TO POSITION THE REST OF THE CAGE ACCORDINGLY. THERE WAS NOT A SIGNIFICANT DELAY TO THE CASE, OR ANY HARM DONE TO THE PATIENT BECAUSE OF THE INCIDENT. ALL THE BROKEN PIECES OF THE IMPLANT WERE REMOVED FROM THE PATIENT AND DISCARDED BY THE SURGICAL STAFF AND CAN'T BE SENT BACK FOR ANALYSIS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 391039 | IDYS-TLIF TIVAC | IDYS TLIF TIVAC CAGE L29MM H09MM, | OVD | CLARIANCE SAS | N1A9X-N113Y | 03700780628643 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown |