FDA Adverse Event Malfunction Summary report: N

IDYS-TLIF TIVAC

MDR report key: 17746225 · Received September 14, 2023

Report

Report Number
3009962553-2023-00003
Event Type
Malfunction
Date Received
September 14, 2023
Date of Event
August 15, 2023
Report Date
September 13, 2023
Manufacturer
CLARIANCE SAS
Product Code
OVD
UDI-DI
03700780628643
PMA / PMN Number
K183259
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MO, US
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

WHILE IMPACTING THE TLIF CAGE INTO THE DISC SPACE, BEFORE ARTICULATING THE INSERTER, THE PROXIMAL TANTALUM MARKER PORTION OF THE CAGE BROKE FROM THE REST OF THE CAGE. THE SURGEON WAS ABLE TO RETRIEVE THE BROKEN PORTION AND REMOVE IT FROM THE DISC SPACE. THEN, THE SURGEON WAS ABLE TO POSITION THE REST OF THE CAGE ACCORDINGLY. THIS EVENT IS REPORTABLE AS A MALFUNCTION.

Additional Manufacturer Narrative · 0

WHILE IMPACTING THE TLIF CAGE INTO THE DISC SPACE, BEFORE ARTICULATING THE INSERTER, THE PROXIMAL TANTALUM MARKER PORTION OF THE CAGE BROKE FROM THE REST OF THE CAGE. THE SURGEON WAS ABLE TO RETRIEVE THE BROKEN PORTION AND REMOVE IT FROM THE DISC SPACE. THEN, THE SURGEON WAS ABLE TO POSITION THE REST OF THE CAGE ACCORDINGLY. THIS EVENT IS REPORTABLE AS A MALFUNCTION. ON 11/28/2023, THIS MEDWATCH 3500A FILING NOW INCLUDES THE ANALYSIS REPORT FROM THE MANUFACTURER.

Description of Event or Problem · 0

THE SURGEON WAS ON THE PATIENT'S RIGHT AT L4-5 USING A 9MM TI COATED PEEK BANANA CAGE WITH THE 25-DEGREE ANGLED INSERTER. WHILE IMPACTING INTO THE DISC SPACE, BEFORE ARTICULATING THE INSERTER, THE PROXIMAL TANTALUM MARKER PORTION OF THE CAGE BROKE FROM THE REST OF THE CAGE. HE WAS ABLE TO RETRIEVE THE BROKEN PORTION AND REMOVE IT FROM THE DISC SPACE. HE THEN USED A TAMP/PUSHER TO POSITION THE REST OF THE CAGE ACCORDINGLY. THERE WAS NOT A SIGNIFICANT DELAY TO THE CASE, OR ANY HARM DONE TO THE PATIENT BECAUSE OF THE INCIDENT. ALL THE BROKEN PIECES OF THE IMPLANT WERE REMOVED FROM THE PATIENT AND DISCARDED BY THE SURGICAL STAFF AND CAN'T BE SENT BACK FOR ANALYSIS.

Description of Event or Problem · 0

THE SURGEON WAS ON THE PATIENT'S RIGHT AT L4-5 USING A 9MM TI COATED PEEK BANANA CAGE WITH THE 25-DEGREE ANGLED INSERTER. WHILE IMPACTING INTO THE DISC SPACE, BEFORE ARTICULATING THE INSERTER, THE PROXIMAL TANTALUM MARKER PORTION OF THE CAGE BROKE FROM THE REST OF THE CAGE. HE WAS ABLE TO RETRIEVE THE BROKEN PORTION AND REMOVE IT FROM THE DISC SPACE. HE THEN USED A TAMP/PUSHER TO POSITION THE REST OF THE CAGE ACCORDINGLY. THERE WAS NOT A SIGNIFICANT DELAY TO THE CASE, OR ANY HARM DONE TO THE PATIENT BECAUSE OF THE INCIDENT. ALL THE BROKEN PIECES OF THE IMPLANT WERE REMOVED FROM THE PATIENT AND DISCARDED BY THE SURGICAL STAFF AND CAN'T BE SENT BACK FOR ANALYSIS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
391039 IDYS-TLIF TIVAC IDYS TLIF TIVAC CAGE L29MM H09MM, OVD CLARIANCE SAS N1A9X-N113Y 03700780628643

Patients

Seq Age Sex Outcome Treatment
1 Unknown