BINAXNOW COVID-19 AG SELF TEST 2CT
Report
- Report Number
- 1221359-2023-01416
- Event Type
- Malfunction
- Date Received
- September 14, 2023
- Date of Event
- August 16, 2023
- Report Date
- September 26, 2023
- Manufacturer
- ABBOTT DIAGNOSTICS SCARBOROUGH, INC.
- Product Code
- QKP
- UDI-DI
- 00811877011408
- PMA / PMN Number
- EUA210264
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- US
- Reporter Occupation
- 003
Narratives
D4-UDI: (B)(4). THE REMAINDER OF THE INVESTIGATION REMAINS IN PROGRESS. A SUPPLEMENTAL REPORT WILL BE PROVIDED AFTER COMPLETION. H3 OTHER TEXT: SINGLE USE; DEVICE DISCARDED.
D4-UDI: (B)(4). TESTING WAS PERFORMED AT ABBOTT DIAGNOSTICS SCARBOROUGH, INC. ON RETAINED KIT LOT 184516 WITH INTERNAL POSITIVE QUALITY CONTROL SAMPLES AND NEGATIVE QUALITY CONTROL SWABS. ALL TEST RESULTS WERE VALID AND PERFORMED AS EXPECTED. ADDITIONALLY, THE MANUFACTURING RECORDS AND QUALITY CONTROL RELEASE TESTING WAS REVIEWED FOR TEST KIT PART NUMBER 195-160/ LOT: 184516, TEST BASE PART NUMBER 195-430WL/ LOT: 181576. THE LOT MET THE REQUIRED RELEASE SPECIFICATIONS. A REVIEW OF THE COMPLAINTS REPORTED AS FALSE NEGATIVE PATIENT RESULTS (CONFIRMED AND UNCONFIRMED, CONFLICTING RESULTS) RELATED TO KIT LOT 184516 SHOWED THAT THE COMPLAINT RATE IS (B)(4). ABBOTT DIAGNOSTICS SCARBOROUGH, INC. WAS UNABLE TO DETERMINE THE EXACT ROOT CAUSE OF THE REPORTED ISSUE; HOWEVER, IT COULD POSSIBLY BE RELATED TO ISSUES INCLUDING THE SELF-TEST USER PERFORMANCE, INTERPRETATION OF THE RESULT, OR THE SPECIFIC PATIENT SAMPLE. H3 OTHER TEXT : SINGLE USE; DEVICE DISCARDED.
THE CONSUMER REPORTED FALSE NEGATIVE RESULTS WITH THE BINAXNOW COVID-19 ANTIGEN SELF-TEST FOR THREE (3) TESTS PERFORMED ON VARIOUS DATES ON A NASAL KITTED SWAB. THIS MFR. REPORT ADDRESSES TEST THREE (3) OF THREE (3). THE CONSUMER REPORTED A FALSE NEGATIVE RESULT WITH THE BINAXNOW COVID-19 ANTIGEN SELF-TEST PERFORMED ON (B)(6) 2023 ON A NASAL KITTED SWAB. CONFIRMATION TESTING WAS NOT PERFORMED. THE CONSUMER STATED THEY WERE SYMPTOMATIC. NO ADDITIONAL PATIENT INFORMATION, INCLUDING TREATMENT AND OUTCOME, WAS PROVIDED.
THE CONSUMER REPORTED FALSE NEGATIVE RESULTS WITH THE BINAXNOW COVID-19 ANTIGEN SELF-TEST FOR THREE (3) TESTS PERFORMED ON VARIOUS DATES ON A NASAL KITTED SWAB. THIS MFR. REPORT ADDRESSES TEST THREE (3) OF THREE (3). THE CONSUMER REPORTED A FALSE NEGATIVE RESULT WITH THE BINAXNOW COVID-19 ANTIGEN SELF-TEST PERFORMED ON (B)(6) 2023 ON A NASAL KITTED SWAB. CONFIRMATION TESTING WAS NOT PERFORMED. THE CONSUMER STATED THEY WERE SYMPTOMATIC. NO ADDITIONAL PATIENT INFORMATION, INCLUDING TREATMENT AND OUTCOME, WAS PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 2183311 | BINAXNOW COVID-19 AG SELF TEST 2CT | CORONAVIRUS ANTIGEN DETECTION TEST SYSTEM. | QKP | ABBOTT DIAGNOSTICS SCARBOROUGH, INC. | 184516 | 00811877011408 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown |