HEATER-COOLER SYSTEM 3T
Report
- Report Number
- 9611109-2023-00430
- Event Type
- Injury
- Date Received
- September 14, 2023
- Report Date
- March 4, 2024
- Manufacturer
- LIVANOVA DEUTSCHLAND
- Product Code
- DWC
- Removal / Correction Number
- Z-2076/2081-2015
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- UK
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
A.1.-A.5. PATIENT INFORMATION WAS NOT PROVIDED. D.4. SERIAL NUMBER IS UNKNOWN. THIS INFORMATION WILL BE PROVIDED IN A SUPPLEMENTAL REPORT IF MADE AVAILABLE. H.4. AS THE SERIAL NUMBER IS UNKNOWN, THE DEVICE MANUFACTURE DATE COULD NOT BE DETERMINED. THIS INFORMATION WILL BE PROVIDED IN A SUPPLEMENTAL REPORT IF MADE AVAILABLE. G.5. THE HEATER-COOLER 16-02-80 IS NOT DISTRIBUTED IN THE USA AND IT IS SIMILAR TO HEATER-COOLER 16-02-85, WHICH IS DISTRIBUTED IN THE USA (510(K) NUMBER: K191402). H.9. LIVANOVA DEUTSCHLAND IMPLEMENTED A FIELD SAFETY NOTICE FOR DISINFECTION AND CLEANING OF HEATER-COOLER DEVICES. THE Z NUMBER IS Z-2076/2081-2015. H10: LIVANOVA DEUTSCHLAND MANUFACTURES THE HEATER-COOLER SYSTEM 3T. THE INCIDENT OCCURRED IN (B)(6). LIVANOVA INITIATED AN INVESTIGATION. IF ANY ADDITIONAL INFORMATION PERTINENT TO THE REPORTED EVENT IS RECEIVED, IT WILL BE PROVIDED IN A SUPPLEMENTAL REPORT.
H10: THROUGH FOLLOW UP COMMUNICATION, LIVANOVA LEARNED THAT POSSIBLE INVOLVED DEVICE SERIAL NUMBERS (MODEL: 16-02-80) COULD BE THE FOLLOWING: (B)(6) (MANUFACTURED 06/APR/11) OR (B)(6) (MANUFACTURED 06/APR/11) OR (B)(6) (MANUFACTURED 06/APR/11) OR (B)(6) (MANUFACTURED 15/NOV/13). NO OTHER INFORMATION HAS BEEN MADE AVAILABLE FROM THE CUSTOMER. DEVICE HISTORY RECORDS (DHRS) REVIEW OF POSSIBLE INVOLVED DEVICE SERIAL NUMBERS HAS BEEN COMPLETED AND DID NOT IDENTIFY ANY DEVIATION OR NON-CONFORMITY RELEVANT TO THE REPORTED ISSUE. NO FURTHER INVESTIGATION IS POSSIBLE. SOURCE OF PATIENT CONTAMINATION REMAINS UNKNOWN AND A DIRECT RELATIONSHIP BETWEEN THE REPORTED ADVERSE EVENT AND THE DEVICE COULD NOT BE ESTABLISHED. LIVANOVA ALREADY IMPLEMENTED A STRATEGY TO DECREASE THE PROBABILITY OF BACTERIA GROW IN THE HC DEVICE BY APPLYING MULTIPLE MEASURES IMPLEMENTED OVER THE PAST FEW YEARS THROUGH DEDICATED CAPA AND A FIELD ACTION. NO DEVICE, BETWEEN THE ONES POSSIBLE INVOLVED AND IN USE AT THE HOSPITAL AT THE TIME OF SURGERIES ((B)(6) 2014), WAS UPGRADED WITH VACUUM AND SEALING KIT. THE RISK IS IN THE ACCEPTABLE REGION. NO SPECIFIC ACTION WAS CURRENTLY DEEMED NECESSARY. LIVANOVA WILL KEEP MONITORING THE MARKET.
THE EVENT WAS IDENTIFIED THROUGH A RETROSPECTIVE REVIEW OF PRODUCT LIABILITY LAWSUITS. THROUGH THIS REVIEW, WE HAVE IDENTIFIED A FEW EVENTS WHERE THE PLAINTIFF HAD FILED A SUIT AGAINST THE COMPANY, AND IT WAS HANDLED BY OUR LEGAL DEPARTMENT, BUT THE UNDERLYING PRODUCT INFORMATION HAD NOT BEEN FORWARDED TO THE COMPLAINT HANDLING UNIT. LIVANOVA OPENED A CAPA TO IDENTIFY ANY POSSIBLE LAWSUITS THAT MIGHT REQUIRE COMPLAINT REPORTING AND TO RETROSPECTIVELY SUBMIT THOSE EVENTS THAT WERE NOT PREVIOUSLY SUBMITTED AND TO PREVENT FUTURE SIMILAR ISSUES FROM OCCURRING. COMPLAINANT ALLEGES THROUGH THEIR COUNSEL A NON TUBERCULOSIS MYCOBACTERIUM INFECTION FOLLOWING A SURGERY UNDERWENT IN (B)(6) 2014 (DATE UNKNOWN). BASED ON CURRENT STATUS OF THE INVESTIGATION THE ALLEGED DEVICE ISSUE WAS YET NOT CONFIRMED.
SEE INITIAL REPORT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 390818 | HEATER-COOLER SYSTEM 3T | CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS | DWC | LIVANOVA DEUTSCHLAND | 16-02-80 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown | Other |