HARMONIC ACE PLUS 7 ADV HEMOSTASIS 36
Report
- Report Number
- 3005075853-2023-06614
- Event Type
- Malfunction
- Date Received
- September 13, 2023
- Date of Event
- August 2, 2023
- Report Date
- September 13, 2023
- Manufacturer
- ETHICON ENDO-SURGERY, LLC.
- Product Code
- LFL
- UDI-DI
- 10705036014621
- PMA / PMN Number
- K132612
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CH
- Reporter Occupation
- OTHER
Narratives
(B)(4). DATE SENT: 9/13/2023. D4 BATCH #: X9679Z. INVESTIGATION SUMMARY: THE PRODUCT WAS RETURNED FOR EVALUATION. VISUAL INSPECTION AND FUNCTIONAL TESTING WERE CONDUCTED ON THE RETURNED DEVICE. VISUAL ANALYSIS OF THE RETURNED SAMPLE DETERMINED THAT THE DEVICE WAS RETURNED WITH THE TISSUE DETACHED NOT RETURNED. IN ADDITION THE BLADE TIP WAS DAMAGED, HOWEVER, THE BLADE WAS NOT CRACKED. THE DEVICE WAS CONNECTED TO A GEN11 AND THE DEVICE DID ACTIVATE DURING FUNCTIONAL TESTING. THE DEVICE WAS DISASSEMBLED TO VERIFY THE CONDITION OF THE INTERNAL COMPONENTS AND NO ANOMALIES WERE FOUND. BASED ON THE CONDITION OF THE TISSUE PAD, A PROBABLE CAUSE FOR THIS DAMAGE IS THAT THE CLAMP OF THE DEVICE MAY HAVE BEEN CLOSED AND THE INSTRUMENT ACTIVATED WITHOUT TISSUE PRESENT. CARE SHOULD BE TAKEN NOT TO APPLY PRESSURE BETWEEN THE INSTRUMENT BLADE AND TISSUE PAD WITHOUT HAVING TISSUE BETWEEN THEM. KEEP THE CLAMP ARM OPEN WHEN BACK CUTTING OR WHILE THE BLADE IS ACTIVE WITHOUT TISSUE BETWEEN THE BLADE AND TISSUE PAD TO AVOID DAMAGE TO THE TISSUE PAD. THE RESULTING DAMAGE CONTRIBUTES TO THE REMOVAL OF THE PAD FROM THE CLAMP ARM. THE CLEANING OF THE PAD, NOT IN ACCORDANCE WITH THE IFU, CAN ALSO RESULT IN THE REMOVAL OF THE PAD DURING USE. PROBABLE CAUSES OF THE BLADE DAMAGE ARE APPLYING PRESSURE BETWEEN THE INSTRUMENT BLADE AND TISSUE PAD WITHOUT HAVING TISSUE BETWEEN THEM, SUBSEQUENT ACTIVATIONS WOULD COMPLETELY WEAR AND MELT THE TISSUE PAD UNTIL THE BLADE TIP MAKES CONTACT WITH THE CLAMP ARM. AVOID ACTIVATING THE BLADE WITHOUT TISSUE BETWEEN THE BLADE AND TISSUE PAD TO PREVENT THIS TYPE OF DAMAGE. AS PART OF THE QUALITY PROCESS, ALL DEVICES ARE MANUFACTURED, INSPECTED, AND RELEASED TO APPROVED SPECIFICATIONS. A MANUFACTURING RECORD EVALUATION WAS PERFORMED FOR THE FINISHED DEVICE LOT/BATCH NUMBER X9679Z, AND NO NON-CONFORMANCES WERE IDENTIFIED.
IT WAS REPORTED THAT DURING A THYROID PROCEDURE THE ¿TIGHTEN ASSEMBLY¿ ALERT SCREEN WAS DISPLAYED BY THE GENERATOR. CHANGED TO ANOTHER DEVICE TO COMPLETE SURGERY. THERE WAS NO PATIENT CONSEQUENCE REPORTED. NO ADDITIONAL INFORMATION CAN BE PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 77761 | HARMONIC ACE PLUS 7 ADV HEMOSTASIS 36 | INSTRUMENT, ULTRASONIC SURGICAL | LFL | ETHICON ENDO-SURGERY, LLC. | X9679Z | 10705036014621 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown | GENERATOR AND HANDPIECE |