ALINITY I TOTAL B-HCG REAGENT KIT
Report
- Report Number
- 3005094123-2023-00240
- Event Type
- Malfunction
- Date Received
- September 13, 2023
- Date of Event
- August 25, 2023
- Report Date
- September 26, 2023
- Manufacturer
- ABBOTT IRELAND DIAGNOSTICS DIVISION
- Product Code
- DHA
- PMA / PMN Number
- K170317
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AS
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
AN EVALUATION IS IN PROCESS. A FOLLOW-UP REPORT WILL BE SUBMITTED WHEN THE EVALUATION IS COMPLETE.
AFTER FURTHER EVALUATION, THE SUSPECT MEDICAL DEVICE WAS CHANGED FROM ALINITY I TOTAL BHCG, LIST 07P51-30, ABBOTT IRELAND DIAGNOSTICS DIVISTION, LONGFORD, IRELAND TO THE ALINITY I PROCESSING MODULE, LIST 03R65-01, ABBOTT LABORATORIES, IRVING, TEXAS, USA. MDR NUMBER 3016438761-2023-00514-00 HAS BEEN SUBMITTED AND ALL FURTHER INFORMATION WILL BE DOCUMENTED UNDER THAT MDR NUMBER.
THE CUSTOMER OBSERVED FALSELY INCREASED ALINITY I TOTAL HCG RESULTS ON ONE PATIENT THAT IS UNLIKELY PREGNANT DUE TO HORMONE PROFILE. THE FOLLOWING DATA WAS PROVIDED: SID (B)(6) PROCESSED ON (B)(6) 2023 INITIAL RESULT = 76. THERE IS NO RETEST RESULT AVAILABLE. SAMPLE WAS A SHORT SAMPLE THAT WAS HEMOLYZED. NO IMPACT TO PATIENT MANAGEMENT WAS REPORTED.
THE CUSTOMER OBSERVED FALSELY INCREASED ALINITY I TOTAL -HCG RESULTS ON ONE PATIENT THAT IS UNLIKELY PREGNANT DUE TO HORMONE PROFILE. THE FOLLOWING DATA WAS PROVIDED: SID (B)(6) PROCESSED ON 25AUG2023 INITIAL RESULT = 76 THERE IS NO RETEST RESULT AVAILABLE SAMPLE WAS A SHORT SAMPLE THAT WAS HEMOLYZED NO IMPACT TO PATIENT MANAGEMENT WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1589724 | ALINITY I TOTAL B-HCG REAGENT KIT | SYSTEM, TEST, HUMAN CHORIONIC GONADOTROPIN | DHA | ABBOTT IRELAND DIAGNOSTICS DIVISION | 45628UD00 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown | ALNTY I PROCESSING MODU, 03R65-01, AI21710| ALNTY I PROCESSING MODU, 03R65-01, AI21710 |