FDA Adverse Event Injury Summary report: N

MIC-KEY GASTROSTOMY FEEDING TUBE KIT (ENFIT)

MDR report key: 17738293 · Received September 13, 2023

Report

Report Number
9611594-2023-00134
Event Type
Injury
Date Received
September 13, 2023
Date of Event
August 25, 2023
Report Date
November 2, 2023
Manufacturer
AVANOS MEDICAL INC.
Product Code
KNT
UDI-DI
00350770427881
PMA / PMN Number
K043114
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

THE ACTUAL COMPLAINT PRODUCT WAS NOT RETURNED FOR EVALUATION. A REVIEW OF THE DEVICE HISTORY RECORD IS IN-PROGRESS. ALL INFORMATION REASONABLY KNOWN AS OF 13 SEP 2023 HAS BEEN INCLUDED IN THIS HEALTH AUTHORITY REPORT. SHOULD ADDITIONAL INFORMATION BE OBTAINED, A FOLLOW-UP HEALTH AUTHORITY REPORT WILL BE PROVIDED. THE INFORMATION PROVIDED BY AVANOS MEDICAL INC. REPRESENTS ALL OF THE KNOWN INFORMATION AT THIS TIME. DESPITE GOOD FAITH EFFORTS TO OBTAIN ADDITIONAL INFORMATION, THE COMPLAINANT/REPORTER WAS UNABLE OR UNWILLING TO PROVIDE ANY FURTHER PATIENT, PRODUCT, OR PROCEDURAL DETAILS TO AVANOS MEDICAL INC. AVANOS MEDICAL INC. HAS NO INDEPENDENT KNOWLEDGE OF THE REPORTED EVENT BUT IS RELAYING THE INFORMATION PROVIDED BY THE USER FACILITY WHERE THE INCIDENT OCCURRED. THIS PRODUCT INCIDENT IS DOCUMENTED IN THE AVANOS MEDICAL COMPLAINT DATABASE AND IDENTIFIED AS COMPLAINT (B)(4). THIS INFORMATION IS SUBMITTED PURSUANT TO 21CFR803, IN COMPLIANCE WITH THE MEDICAL DEVICE REPORTING REQUIREMENT AND SHOULD NOT BE CONSIDERED TO BE AN ADMISSION THAT AN AVANOS MEDICAL INC. PRODUCT IS DEFECTIVE OR CAUSED SERIOUS INJURY. H3 OTHER TEXT : DEVICE NOT RETURNED

Additional Manufacturer Narrative · 0

CORRECTION: D9. THE PRODUCT INVOLVED IN THE REPORT HAS BEEN RETURNED AND THE INVESTIGATION REMAINS IN PROGRESS AT THIS TIME. ALL INFORMATION REASONABLY KNOWN AS OF 10 OCT 2023 HAS BEEN INCLUDED IN THIS HEALTH AUTHORITY REPORT. SHOULD ADDITIONAL INFORMATION BE OBTAINED, A FOLLOW-UP HEALTH AUTHORITY REPORT WILL BE PROVIDED. THE INFORMATION PROVIDED BY AVANOS MEDICAL INC. REPRESENTS ALL OF THE KNOWN INFORMATION AT THIS TIME. DESPITE GOOD FAITH EFFORTS TO OBTAIN ADDITIONAL INFORMATION, THE COMPLAINANT/REPORTER WAS UNABLE OR UNWILLING TO PROVIDE ANY FURTHER PATIENT, PRODUCT, OR PROCEDURAL DETAILS TO AVANOS MEDICAL INC. AVANOS MEDICAL INC. HAS NO INDEPENDENT KNOWLEDGE OF THE REPORTED EVENT BUT IS RELAYING THE INFORMATION PROVIDED BY THE USER FACILITY WHERE THE INCIDENT OCCURRED. THIS PRODUCT INCIDENT IS DOCUMENTED IN THE AVANOS MEDICAL COMPLAINT DATABASE AND IDENTIFIED AS COMPLAINT (B)(4). THIS INFORMATION IS SUBMITTED PURSUANT TO 21CFR803, IN COMPLIANCE WITH THE MEDICAL DEVICE REPORTING REQUIREMENT AND SHOULD NOT BE CONSIDERED TO BE AN ADMISSION THAT AN AVANOS MEDICAL INC. PRODUCT IS DEFECTIVE OR CAUSED SERIOUS INJURY.

Additional Manufacturer Narrative · 0

H6: INVESTIGATION FINDINGS: APPROPRIATE TERM/CODE NOT AVAILABLE: MALFUNCTION OBSERVED WITHOUT CONCLUSIVE FINDING THE DEVICE HISTORY RECORD FOR LOT 30242271 WAS REVIEWED AND THE PRODUCT WAS PRODUCED ACCORDING TO PRODUCT SPECIFICATIONS. THE ACTUAL COMPLAINT PRODUCT WAS RETURNED FOR EVALUATION. THE SAMPLE WAS RECEIVED WITHOUT A SECURE LOCK TIP AT THE DISTAL END OF THE EXTENSION SET TUBING. THE END WAS INSPECTED FOR PRESENCE OF ADHESIVE RESIDUE. ONE SIDE EXHIBITS NO VISIBLE RESIDUE. THE OPPOSITE SIDE EXHIBITS MINIMAL/INCONSISTENT RESIDUE. THE INCIDENT WAS CONFIRMED AS REPORTED. A ROOT CAUSE COULD NOT BE DETERMINED. ALL INFORMATION REASONABLY KNOWN AS OF 31 OCT 2023 HAS BEEN INCLUDED IN THIS HEALTH AUTHORITY REPORT. SHOULD ADDITIONAL INFORMATION BE OBTAINED, A FOLLOW-UP HEALTH AUTHORITY REPORT WILL BE PROVIDED. THE INFORMATION PROVIDED BY AVANOS MEDICAL INC. REPRESENTS ALL OF THE KNOWN INFORMATION AT THIS TIME. DESPITE GOOD FAITH EFFORTS TO OBTAIN ADDITIONAL INFORMATION, THE COMPLAINANT/REPORTER WAS UNABLE OR UNWILLING TO PROVIDE ANY FURTHER PATIENT, PRODUCT, OR PROCEDURAL DETAILS TO AVANOS MEDICAL INC. AVANOS MEDICAL INC. HAS NO INDEPENDENT KNOWLEDGE OF THE REPORTED EVENT BUT IS RELAYING THE INFORMATION PROVIDED BY THE USER FACILITY WHERE THE INCIDENT OCCURRED. THIS PRODUCT INCIDENT IS DOCUMENTED IN THE AVANOS MEDICAL COMPLAINT DATABASE AND IDENTIFIED AS COMPLAINT (B)(4). THIS INFORMATION IS SUBMITTED PURSUANT TO 21CFR803, IN COMPLIANCE WITH THE MEDICAL DEVICE REPORTING REQUIREMENT AND SHOULD NOT BE CONSIDERED TO BE AN ADMISSION THAT AN AVANOS MEDICAL INC. PRODUCT IS DEFECTIVE OR CAUSED SERIOUS INJURY.

Description of Event or Problem · 0

IT WAS REPORTED, THE CONNECTION BETWEEN ENFIT CONNECTOR AND MIC-KEY TUBE DETACHED, AND NUTRITION LEAKED FROM THERE. A PATIENT HAD BEEN TREATED WITH INSULIN. THE PATIENT BECAME HYPOGLYCEMIC AND REQUIRED MEDICAL TREATMENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1789339 MIC-KEY GASTROSTOMY FEEDING TUBE KIT (ENFIT) DH EF BALLOON TUBES PRODUCTS KNT AVANOS MEDICAL INC. 8140-20-3.5 30242271 00350770427881

Patients

Seq Age Sex Outcome Treatment
1 Unknown Other