FDA Adverse Event Injury Summary report: N

AERO CHROME* BREATHABLE PERFORMANCE SURGICAL GOWN WITH TOWEL- XLARGE-STERILE

MDR report key: 17736688 · Received September 13, 2023

Report

Report Number
9680646-2023-00002
Event Type
Injury
Date Received
September 13, 2023
Report Date
November 14, 2023
Manufacturer
O&M HALYARD, INC.
Product Code
KYA
UDI-DI
30680651446740
PMA / PMN Number
K200522
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

THE PRODUCT INVOLVED IN THE EVENT WAS NOT RETURNED. A FOLLOW-UP REPORT WILL BE PROVIDED UPON CONCLUSION OF INVESTIGATION. THIS PRODUCT INCIDENT IS DOCUMENTED IN THE O&M HALYARD, INC. COMPLAINT DATABASE AND IDENTIFIED AS COMPLAINT (B)(4). THIS INFORMATION IS SUBMITTED PURSUANT TO (B)(4), IN COMPLIANCE WITH THE MEDICAL DEVICE REPORTING REQUIREMENT AND SHOULD NOT BE CONSIDERED TO BE AN ADMISSION THAT A O&M HALYARD, INC. PRODUCT IS DEFECTIVE OR HAS CAUSED SERIOUS INJURY.

Additional Manufacturer Narrative · 0

THE PRODUCT INVOLVED IN THE EVENT WAS NOT RETURNED FOR EVALUATION. THE INVESTIGATION WAS PERFORMED PER INFORMATION PROVIDED BY THE CUSTOMER. A SAMPLE, PHOTO OR LOT NUMBER WAS NOT AVAILABLE TO PERFORM AN EVALUATION. WE WERE UNABLE TO REVIEW THE DEVICE HISTORY RECORD FOR THIS COMPLAINT GIVEN THE REPORTER DID NOT PROVIDE A COMPLAINT LOT NUMBER. WE REACHED OUT TO THE CUSTOMER FOR ADDITIONAL INFORMATION ON AUGUST 15, 18, AND 24TH TO OBTAIN ADDITIONAL DETAILS. THE REPORTER HAS NOT RESPONDED. THE SAFETY CLEARANCE ASSESSMENT OF THE AERO BLUE SURGICAL GOWN WAS REVIEWED AND CONFIRMED THAT THE GOWN IS NOT CONSIDERED TO BE AN IRRITANT. BASED ON THE INFORMATION AVAILABLE AND INVESTIGATION PERFORMED, A ROOT CAUSE WAS NOT IDENTIFIED. ACTIONS TAKEN: NOTIFICATION OF THE REPORTED INCIDENT WAS SENT TO THE GOWN'S MANUFACTURING AREA. COMPLAINT DATA IS MONITORED TO IDENTIFY UPWARD TRENDS ASSOCIATED WITH THE REPORTED INCIDENT/FAMILY. A REVIEW OF THE MANUFACTURING AREA'S CLEANING PROCESS WAS PERFORMED AND THERE HAVE BEEN NO CHANGES. PER PROCEDURE, THE WORKING STATIONS ARE CLEANED AT THE BEGINNING OF EACH SHIFT. THE PRODUCT INCIDENT IS DOCUMENTED IN THE O&M HALYARD, INC. COMPLAINT DATABASE AND IDENTIFIED AS COMPLAINT (B)(4). THIS INFORMATION IS SUBMITTED PURSUANT TO 21CFR803, IN COMPLIANCE WITH THE MEDICAL DEVICE REPORTING REQUIREMENT AND SHOULD NOT BE CONSIDERED TO BE AN ADMISSION THAT AN O&M HALYARD, INC. PRODUCT IS DEFECTIVE OR HAS CAUSED SERIOUS INJURY. H3 OTHER TEXT: DEVICE NOT RETURNED.

Description of Event or Problem · 0

DERMATOLOGIST REPORTS THAT HIS PATIENT HAS A SUSPECTED ALLERGY TO AN INGREDIENT IN SLEEVE AND COLLAR PORTION OF SURGICAL GOWNS. THE PATIENT UTILIZES AERO BLUE PERFORMANCE SURGICAL GOWN (REF # (B)(4)) AND THE AERO CHROME BREATHABLE PERFORMANCE SURGICAL GOWN (REF # (B)(4)). PATIENT HAS A HISTORY OF ALLERGIC CONTACT DERMATITIS TO SEVERAL ALLERGENS, INCLUDING CARBA MIX IN SURGICAL GLOVES. PATIENT PRESENTS WITH RASH ON HANDS AND CENTRAL NECK, THAT WORSENS AFTER SURGICAL DAYS. THESE AREAS ARE TOUCHED BY THE CUFF AND NECKLINE OF THE SURGICAL GOWNS. PATIENT RECEIVES TOPICAL PRESCRIPTION STEROIDS AND OTHER TEATMENTS NOT SPECIFIED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
89066 AERO CHROME* BREATHABLE PERFORMANCE SURGICAL GOWN WITH TOWEL- XLARGE-STERILE SURGICAL GOWNS KYA O&M HALYARD, INC. 44674 UNKNOWN 30680651446740
906806 AERO CHROME* BREATHABLE PERFORMANCE SURGICAL GOWN WITH TOWEL- XLARGE-STERILE SURGICAL GOWNS KYA O&M HALYARD, INC. 44674 UNKNOWN 30680651446740

Patients

Seq Age Sex Outcome Treatment
1 Male Other