FDA Adverse Event Malfunction Summary report: N

ALLURA XPER FD

MDR report key: 17734621 · Received September 13, 2023

Report

Report Number
3003768277-2023-04980
Event Type
Malfunction
Date Received
September 13, 2023
Date of Event
August 14, 2023
Manufacturer
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
Product Code
OWB
UDI-DI
00884838054196
PMA / PMN Number
K161563
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

PHILIPS HAS INVESTIGATED THIS COMPLAINT. ACCORDING TO THE INFORMATION COLLECTED, THE SYSTEM WAS OUTSIDE OF CLINICAL USE AT THE TIME OF THE EVENT. THE PHILIPS FIELD SERVICE ENGINEER (FSE) INSPECTED THE SYSTEM ONSITE AND CONFIRMED THAT THERE WAS A DISK SPACE ISSUE ON THE ALLURA XPER FD10/10 SYSTEM AND THEREFORE IT WOULD PREVENT IMAGE PROCESSING. REVIEW OF LOG FILE SHOWED THAT MALFUNCTIONING FRONTAL IP-PC. FSE REPLACED IPPC AND DATA DRIVES. AFTER REPLACEMENT THE SYSTEM WAS RETURNED TO USE IN GOOD WORKING ORDER. THE FAILURE ANALYSIS WAS EXECUTED HDD BAY/LABEL WAS REPLACED PER PRQ. BAY LIGHTS NEEDED THEIR INDIVIDUAL POWER BUTTONS PRESSED TO COME ON; ISSUE WAS RESOLVED AFTER THIS WAS DONE. CMOS BATTERY CHANGED DUE TO LOW CHARGE. FINAL TESTING COMPLETED AND PASSED. THE CODES WERE UPDATED BASED ON THE INVESTIGATION OUTCOME.

Description of Event or Problem · 0

IT WAS REPORTED TO PHILIPS THAT THERE WAS A DISK SPACE ISSUE ON THE ALLURA XPER FD10/10 SYSTEM AND THEREFORE IT WOULD PREVENT IMAGE PROCESSING. NO HARM WAS REPORTED TO PHILIPS. A PHILIPS FIELD SERVICE ENGINEER (FSE) VISITED THE SITE AND REPLACED THE CLARITYIQ II IMAGE PROCESSING PC NO HDD AND THE HARD DISKS. THE DEVICE WAS RETURNED TO EXPECTED FUNCTIONALITY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
676683 ALLURA XPER FD INTERVENTIONAL FLUOROSCOPIC X-RAY SYSTEM OWB PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. ALLURA XPER FD10/10 00884838054196

Patients

Seq Age Sex Outcome Treatment
1 Unknown