FDA Adverse Event Injury Summary report: N

TALENT CAPTIVIA STENT GRAFT SYSTEM

MDR report key: 1772402 · Received July 23, 2010

Report

Report Number
2953200-2010-01386
Event Type
Injury
Date Received
July 23, 2010
Date of Event
June 24, 2010
Report Date
June 24, 2010
Manufacturer
MEDTRONIC CARDIOVASCULAR GALWAY
Product Code
MIH
PMA / PMN Number
P070007
Removal / Correction Number
NA
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

(B)(4): EVALUATION RESULTS: (ARTERIAL AND VESSEL OCCLUSION); (SIGNIFICANT TORTUOUS AORTIC ARCH, MODERATE AMOUNT OF THROMBUS); (PARTIAL COVERAGE OF THE SUBCLAVIAN ARTERY). CONCLUSION: (SIGNIFICANT TORTUOUS AORTIC ARCH, MODERATE AMOUNT OF THROMBUS); (PARTIAL COVERAGE OF THE SUBCLAVIAN ARTERY).

Description of Event or Problem · 1

A TALENT CAPTIVIA STENT GRAFT SYSTEM WAS IMPLANTED IN A PT FOR THE ENDOVASCULAR TREATMENT OF 6.6 CM THORACIC AORTIC ANEURYSM. THE AORTIC ARCH HAD SIGNIFICANT TORTUOSITY, AND THERE WAS A MODERATE AMOUNT OF THROMBUS DISTAL TO THE ANEURYSM. ILIAC VESSEL MORPHOLOGY WAS NOT REPORTED. IT WAS REPORTED THAT WHEN DEPLOYING THE FIRST TALENT CAPTIVIA STENT GRAFT (MFR REPORT # 2953200-2010-01385), THE INJECTOR TUBING DISCONNECTED AND RELEASED CONTRAST EVERYWHERE, INCLUDING ONTO THE PHYSICIAN. CONSEQUENTLY, THE PHYSICIAN INADVERTENTLY PULLED THE STENT GRAFT DOWNWARD APPROX 2 CM, ALTHOUGH THE FINAL POSITION OF THE GRAFT ENDED UP 1 CM AWAY FROM THE SUBCLAVIAN ARTERY. A DISTAL EXTENSION STENT GRAFT WAS THEN DEPLOYED AND BALLOONED WITHOUT ISSUES. ON THE COMPLETION ANGIOGRAM, A PROXIMAL TYPE I ENDOLEAK WAS OBSERVED. THE PHYSICIAN ELECTED TO ADD A PROXIMAL EXTENSION STENT GRAFT, WHICH SUCCESSFULLY RESOLVED THE ENDOLEAK BUT PARTIALLY COVERED THE SUBCLAVIAN ARTERY IN ORDER TO GAIN A PROXIMAL SEAL. NO INTERVENTION WAS PERFORMED FOR THE PARTIAL VESSEL COVERAGE. NO ADDITIONAL CLINICAL SEQUELAE WERE REPORTED, AND THE PT IS FINE. PLEASE NOTE THAT THIS DEVICE THE TALENT CAPTIVIA STENT GRAFT SYSTEM IS SIMILAR TO THE UNITED STATES DISTRIBUTED PRODUCT TALENT THORACIC STENT GRAFT SYSTEM.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 TALENT CAPTIVIA STENT GRAFT SYSTEM MIH MEDTRONIC CARDIOVASCULAR GALWAY NA V00417968

Patients

Seq Age Sex Outcome Treatment
1 UNK