OLYMPUS EVIS GASROINTESTINAL VIDEOSCOPE
Report
- Report Number
- 8010047-2010-00142
- Event Type
- Other
- Date Received
- July 22, 2010
- Report Date
- June 23, 2010
- Manufacturer
- OLYMPUS MEDICAL SYSTEMS CORPORATION
- Product Code
- FDS
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MN, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
THE SUBJECT DEVICE WAS NOT RETURNED TO OLYMPUS FOR EVALUATION. AN OLYMPUS ENDOSCOPE SERVICE SPECIALIST (ESS) HAS VISITED THE FACILITY AND PROVIDED INSERVICING REGARDING THE APPROPRIATE REPROCESSING OF THE SUBJECT DEVICES. THE USER FACILITY'S INFECTION CONTROL DEPARTMENT HAS BEEN NOTIFIED AND IS REVIEWING THE MATTER. BASED UPON THE INFORMATION PROVIDED, THE USER FACILITY DID NOT REPROCESS THE DEVICE IN ACCORDANCE WITH THE INSTRUCTIONS FOR USE.
OLYMPUS WAS INFORMED THAT THE USER FACILITY WAS NOT REPROCESSING 2-CHANNEL GASTROSCOPES IN ACCORDANCE WITH THE DIRECTIONS FOR USE. THE USER FACILITY WAS REPORTED TO BE PERFORMING MANUAL REPROCESSING PRIOR TO PLACING THE DEVICES IN AN AUTOMATIC ENDOSCOPE REPROCESSOR (AER), HOWEVER, THEY WERE NOT REMOVING THE SUCTION SWITCHOVER LEVER, AND WERE NOT BRUSHING THE INSTRUMENT CHANNEL CONNECTING PIPE. THE PRESENCE OF THE SUCTION SWITCHOVER LEVER IN THE DEVICE WHILE THE ENDOSCOPE WAS IN THE AER COULD POTENTIALLY RESULT IN INADEQUATE FLUSHING OF THE INSTRUMENT CHANNELS. THE PRACTICE WAS SAID TO HAVE BEEN ONGOING FOR AN EXTENDED PERIOD OF TIME, POSSIBLY FROM THE TIME OF INITIAL PURCHASES OF THE GASTROSCOPES. THERE WERE NO REPORTS OF ANY PATIENT INFECTIONS ASSOCIATED WITH THIS MATTER.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | OLYMPUS EVIS GASROINTESTINAL VIDEOSCOPE | GASTROINTESTINAL VIDEOSCOPE | FDS | OLYMPUS MEDICAL SYSTEMS CORPORATION | GIF-2T100 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | THE ENDOSCOPES WERE SAID TO BE REPROCESSED IN A| MEDIVATORS DSD-201 AUTOMATIC ENDOSCOPE REPROCESSOR |