TITAN SGS
Report
- Report Number
- 3012481535-2023-00007
- Event Type
- Malfunction
- Date Received
- September 11, 2023
- Date of Event
- July 24, 2023
- Report Date
- August 25, 2023
- Manufacturer
- STANDARD BARIATRICS
- Product Code
- GDW
- UDI-DI
- 00851677007089
- PMA / PMN Number
- K210278
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- KY, US
- Reporter Occupation
- PHYSICIAN
- Health Professional
- Yes
Narratives
UPON EVALUATION OF THE TITAN SGS, IT WAS DETERMINED THAT THE LOAD APPLIED TO THE CLOSURE SCREW WAS EXCESSIVE AND CAUSED IT TO BEND/DISTORT AS WELL AS CAUSING DAMAGE TO THE GEAR DURING THE OPENING OF THE JAWS OF THE TITAN STAPLER, AS DAMAGE WAS OBSERVED. THE INVESTIGATION FOUND THAT THE DAMAGE TO THE CLOSURE SYSTEM LED TO THE DEVICE BEING INOPERABLE. THE ROOT CAUSE WAS ATTRIBUTED TO EXCESSIVE FORCE/ROTATIONS BEING APPLIED BY THE USER TO THE CLOSURE SYSTEM WHILE ATTEMPTING TO MANUALLY OPEN THE JAWS OF THE DEVICE USING THE BAILOUT KEY, RESULTING IN THE DEVICE BEING STUCK IN THE OPEN POSITION.
THE TITAN STAPLER HIT A CRITICAL HIGH LOAD AFTER CONSISTENT HIGH LOAD WARNINGS OCCURED. THE BAILOUT PROCESS WAS INITIATED BY A RESIDENT AND NEARLY COMPLETED SUCCESSFULLY. THE REMAINING PORTION OF THE PROCEDURE WAS COMPLETED WITH AN ETHICON STAPLER. WHEN THE RESIDENT TRIED TO CLOSE THE STAPLER USING THE BAILOUT KEY, THE SILVER KNOB DROPPED OUT OF THE STAPLER. THE SURGEON USED BOVIE AND TOWEL CLAMP TO GET OPEN STAPLER OUT OF ABDOMEN, AS WELL AS INCREASING THE TROCAR INCISION. THE PATIENT IS DOING WELL. A LEAK TEST WAS PERFORMED WITH NO SIGNS OF A LEAK.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1879450 | TITAN SGS | SURGICAL STAPLER WITH STAPLE, IMPLANTABLE, PRODUCT CODE: GDW, PRODUCT CODE: GDW | GDW | STANDARD BARIATRICS | SGS23R | 271-22 | 00851677007089 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Prefer Not To Disclose | Required Intervention |