FDA Adverse Event Malfunction Summary report: N

TITAN SGS

MDR report key: 17720544 · Received September 11, 2023

Report

Report Number
3012481535-2023-00007
Event Type
Malfunction
Date Received
September 11, 2023
Date of Event
July 24, 2023
Report Date
August 25, 2023
Manufacturer
STANDARD BARIATRICS
Product Code
GDW
UDI-DI
00851677007089
PMA / PMN Number
K210278
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
KY, US
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

UPON EVALUATION OF THE TITAN SGS, IT WAS DETERMINED THAT THE LOAD APPLIED TO THE CLOSURE SCREW WAS EXCESSIVE AND CAUSED IT TO BEND/DISTORT AS WELL AS CAUSING DAMAGE TO THE GEAR DURING THE OPENING OF THE JAWS OF THE TITAN STAPLER, AS DAMAGE WAS OBSERVED. THE INVESTIGATION FOUND THAT THE DAMAGE TO THE CLOSURE SYSTEM LED TO THE DEVICE BEING INOPERABLE. THE ROOT CAUSE WAS ATTRIBUTED TO EXCESSIVE FORCE/ROTATIONS BEING APPLIED BY THE USER TO THE CLOSURE SYSTEM WHILE ATTEMPTING TO MANUALLY OPEN THE JAWS OF THE DEVICE USING THE BAILOUT KEY, RESULTING IN THE DEVICE BEING STUCK IN THE OPEN POSITION.

Description of Event or Problem · 0

THE TITAN STAPLER HIT A CRITICAL HIGH LOAD AFTER CONSISTENT HIGH LOAD WARNINGS OCCURED. THE BAILOUT PROCESS WAS INITIATED BY A RESIDENT AND NEARLY COMPLETED SUCCESSFULLY. THE REMAINING PORTION OF THE PROCEDURE WAS COMPLETED WITH AN ETHICON STAPLER. WHEN THE RESIDENT TRIED TO CLOSE THE STAPLER USING THE BAILOUT KEY, THE SILVER KNOB DROPPED OUT OF THE STAPLER. THE SURGEON USED BOVIE AND TOWEL CLAMP TO GET OPEN STAPLER OUT OF ABDOMEN, AS WELL AS INCREASING THE TROCAR INCISION. THE PATIENT IS DOING WELL. A LEAK TEST WAS PERFORMED WITH NO SIGNS OF A LEAK.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1879450 TITAN SGS SURGICAL STAPLER WITH STAPLE, IMPLANTABLE, PRODUCT CODE: GDW, PRODUCT CODE: GDW GDW STANDARD BARIATRICS SGS23R 271-22 00851677007089

Patients

Seq Age Sex Outcome Treatment
1 Prefer Not To Disclose Required Intervention