HARMONIC ACE PLUS 7 ADV HEMOSTASIS 36
Report
- Report Number
- 3005075853-2023-06480
- Event Type
- Malfunction
- Date Received
- September 11, 2023
- Date of Event
- June 30, 2023
- Report Date
- September 11, 2023
- Manufacturer
- ETHICON ENDO-SURGERY, LLC.
- Product Code
- LFL
- UDI-DI
- 10705036014621
- PMA / PMN Number
- K132612
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CH
- Reporter Occupation
- OTHER
Narratives
(B)(4). DATE SENT: 9/11/2023. D4 BATCH #: A9C05A. INVESTIGATION SUMMARY: THE PRODUCT WAS RETURNED FOR EVALUATION. VISUAL INSPECTION AND FUNCTIONAL TESTING WERE CONDUCTED ON THE RETURNED DEVICE. VISUAL ANALYSIS OF THE RETURNED SAMPLE DETERMINED THAT THE DEVICE WAS RETURNED WITH THE DISTAL TIP OF THE BLADE BROKEN OFF AND NOT RETURNED. THE REMAINING BLADE PORTION WAS SCRATCHED, AND WITH EVIDENCE OF CONTACT WITH METAL IN OR OUT OF THE OPERATIVE FIELD. DURING FUNCTIONAL TESTING ON GEN11, AN ALERT SCREEN WAS DISPLAYED. A PROBABLE CAUSE FOR THE DEVICE TO STOP ACTIVATING AND THE GEN11 TO DISPLAY AN ALERT SCREEN IS BLADE DAMAGE. THE DEVICE WAS DISASSEMBLED TO INSPECT THE INTERNAL COMPONENTS AND NO ANOMALIES WERE FOUND. PROBABLE CAUSES OF BLADE DAMAGE, INCLUDING BREAKAGE, ARE EXTERNAL CONTACT DURING PRE-OP OR GENERAL USE, BLADE CONTACT WITH OTHER DEVICES, STAPLES OR CLIPS DURING THE PROCEDURE OR USING ANY MEANS OTHER THAN THE BLADE WRENCH TO ATTACH OR DETACH THE BLADE. ONCE MINOR BLADE DAMAGE HAS OCCURRED, SUBSEQUENT ACTIVATIONS MAY INCREASE THE SEVERITY OF THE BLADE DAMAGE. THIS IN TURN CAN RESULT IN ACTIVATION ISSUES SUCH AS FAILING THE PRE-RUN TEST WITH THE GENERATOR AND DISPLAYING AN ALERT SCREEN. THESE ALERT SCREENS THAT CAN RESULT ARE SUCH AS ¿TIGHTEN ASSEMBLY¿, ¿BLADE ERROR DETECTED¿ OR "RELAXED PRESSURE ON BLADE" FOLLOWED BY A ¿REPLACE INSTRUMENT¿ SCREEN LATER IN THE PROCEDURE. CONTINUED USAGE OF THE DAMAGE BLADE CAN RESULT IN A BROKEN BLADE. AS PART OF THE QUALITY PROCESS ALL DEVICES ARE MANUFACTURED, INSPECTED, AND RELEASED TO APPROVED SPECIFICATIONS. A MANUFACTURING RECORD EVALUATION WAS PERFORMED FOR THE FINISHED DEVICE LOT/BATCH A9C05A AND NO NON-CONFORMANCES WERE IDENTIFIED.
IT WAS REPORTED THAT DURING A LOBECTOMY THE DEVICE WAS HARD TO BE ACTIVATED. SOMETIMES CAN BE ACTIVATED, AND SOMETIMES COULDN'T. CHANGED TO ANOTHER DEVICE TO COMPLETE SURGERY. THERE WAS NO PATIENT CONSEQUENCE REPORTED. NO ADDITIONAL INFORMATION CAN BE PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 2146753 | HARMONIC ACE PLUS 7 ADV HEMOSTASIS 36 | INSTRUMENT, ULTRASONIC SURGICAL | LFL | ETHICON ENDO-SURGERY, LLC. | A9C05A | 10705036014621 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown | GENERATOR AND HANDPIECE |