FDA Adverse Event Malfunction Summary report: N

PENTAX

MDR report key: 17717100 · Received September 10, 2023

Report

Report Number
9610877-2023-58800
Event Type
Malfunction
Date Received
September 10, 2023
Date of Event
August 30, 2023
Report Date
September 11, 2023
Manufacturer
HOYA CORPORATION PENTAX TOKYO OFFICE
Product Code
FDT
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

THIS DEVICE IS CLASSIFIED AS IMPORT FOR EXPORT, THEREFORE 510K IS NOT APPLICABLE. MODEL ED34-I10T-US IS AVAILABLE IN THE USA WITH A 510K NUMBER K163614. THE PRODUCT WAS RETURNED TO PENTAX MEDICAL FOR REPAIR. OUR TECHNICIAN CHECKED THE RETURNED UNIT AND CONFIRMED THAT THE CCD IMAGE DISTORTION. BASED ON THE RESULT, WE CONCLUDED THAT IT WAS CAUSED DUE TO THE EXCESSIVE FORCE APPLIED ON THE CCD MODULE. BASED ON THE TECHNICAL REPORT AND/OR THE RISK ANALYSIS RESULTS, IT WAS EVALUATED TO SUBMIT MDR.

Description of Event or Problem · 0

THE TIME OF EVENT IS UNKNOWN.THERE WAS NO REPORT OF PATIENT HARM.VIDEO IMAGE FAILURE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1891750 PENTAX VIDEO DUODENOSCOPE - I10 STANDARD FDT HOYA CORPORATION PENTAX TOKYO OFFICE ED34-I10T

Patients

Seq Age Sex Outcome Treatment
1 Unknown