FDA Adverse Event Malfunction Summary report: N

TAXUS LIBERTE PACLITAXEL-ELUTING CORONARY STENT SYSTEM

MDR report key: 1771662 · Received July 27, 2010

Report

Report Number
2134265-2010-03142
Event Type
Malfunction
Date Received
July 27, 2010
Date of Event
May 26, 2010
Report Date
June 29, 2010
Manufacturer
BOSTON SCIENTIFIC - GALWAY
Product Code
NIQ
PMA / PMN Number
P060008
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
KR
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

DEVICE EVALUATION: VISUAL AND MICROSCOPIC EXAMINATION REVEALED THAT THE STRUTS ON ROWS 1 TO 5 FROM THE DISTAL EDGE WERE MISALIGNED AND PUSHED 2MM OF THE DISTAL MARKERBAND. THE STRUTS ON ROWS 1 AND 2 FROM THE PROXIMAL EDGE WERE RAISED. THE BALLOON AND TIP SECTIONS OF THE DEVICE WERE VISUALLY AND MICROSCOPICALLY EXAMINED AND NO ISSUES WERE NOTED WITH THEIR PROFILES THAT COULD HAVE POTENTIALLY CONTRIBUTED TO THE COMPLAINT INCIDENT. THE BALLOON WAS TIGHTLY WRAPPED AND WAS NOT SUBJECTED TO ANY POSITIVE PRESSURE. A 0.015 INCH PRODUCT MANDREL WAS INSERTED THROUGH THE LUMEN WITH NO RESTRICTIONS NOTED. THE MANUFACTURING BATCH RECORD REVIEW CONFIRMED THAT THE DEVICE MET ALL MATERIAL, ASSEMBLY AND PERFORMANCE SPECIFICATIONS. THE MOST PROBABLE ROOT CAUSE IS OPERATIONAL CONTEXT AS DEVICE PERFORMANCE WAS LIMITED DUE TO ANATOMICAL AND/OR PROCEDURAL FACTORS. (B)(4).

Description of Event or Problem · 1

REPORTABLE BASED ON THE PRODUCT ANALYSIS COMPLETED ON (B)(6) 2010. IT WAS REPORTED THAT DURING A STENTING TREATMENT PROCEDURE THE DEVICE WAS UNABLE TO CROSS THE LESION. THE LESION WAS LOCATED IN THE CALCIFIED AND SEVERELY TORTUOUS MID LEFT ANTERIOR DESCENDING ARTERY. THE PHYSICIAN ADVANCED THE 2.75X38MM TAXUS LIBERTE STENT TO THE LESION, BUT WAS UNABLE TO CROSS. THERE WERE NO PATIENT COMPLICATIONS. THE PROCEDURE WAS COMPLETED WITH ANOTHER 2.75X38MM TAXUS LIBERTE STENT. PATIENT STATUS IS REPORTED AS STABLE. HOWEVER, THE RETURNED PRODUCT ANALYSIS REVEALED THAT THERE WAS STENT DAMAGE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 TAXUS LIBERTE PACLITAXEL-ELUTING CORONARY STENT SYSTEM CORONARY DRUG-ELUTING STENT NIQ BOSTON SCIENTIFIC - GALWAY H7493894038270 12701867

Patients

Seq Age Sex Outcome Treatment
1 63 YR