FDA Adverse Event Injury Summary report: N

INSIGNIA

MDR report key: 1771631 · Received July 27, 2010

Report

Report Number
2124215-2010-14750
Event Type
Injury
Date Received
July 27, 2010
Date of Event
May 20, 2009
Report Date
October 31, 2012
Manufacturer
GUIDANT CRM CLONMEL IRELAND
Product Code
NVZ
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE

Narratives

Additional Manufacturer Narrative · 1

UPON RECEIPT AT OUR POST MARKET QUALITY ASSURANCE LABORATORY, A THOROUGH EVALUATION OF THE DEVICE WAS PERFORMED. A VISUAL INSPECTION OF THE DEVICE HEADER AND CASE NOTED NO ANOMALIES. THE DEVICE WAS THEN EXPOSED TO SIMULATED HEART LOAD CONDITIONS, AND THE PACING AND SENSING FUNCTIONS WERE TESTED. THE DEVICE OPERATED APPROPRIATELY, ACCORDING TO ITS PERFORMANCE SPECIFICATIONS WITH NO INTERRUPTIONS IN THERAPY OUTPUT OR PROGRAMMER COMMUNICATION AT THE RETURNED PROGRAMMED SETTINGS. A SERIES OF ELECTRICAL TESTS WAS ALSO PERFORMED, AND AGAIN, NORMAL DEVICE FUNCTION WAS OBSERVED.

Additional Manufacturer Narrative · 1

AS OF THIS DATE, THE PACEMAKER HAS NOT BEEN RETURNED. IF THIS PACEMAKER SHOULD BE RETURNED TO OUR COMPANY, IT WILL BE ANALYZED AND THIS EVENT WILL BE REOPENED AND UPDATED AS NECESSARY.

Additional Manufacturer Narrative · 1

AS NO FURTHER INFORMATION CONCERNING THIS REPORT IS EXPECTED, OUR INVESTIGATION IS COMPLETE. THIS INVESTIGATION WILL BE UPDATED SHOULD FURTHER INFORMATION BE PROVIDED.

Description of Event or Problem · 1

BOSTON SCIENTIFIC CRM RECEIVED INFORMATION THAT THIS PACEMAKER BATTERY STATUS INDICATOR HAS FLUCTUATING READINGS. ONE FOLLOW UP VISIT IT STATED BEGINNING OF LIFE (BOL) AND THEN THE NEXT VISIT IT STATED LESS THAN 3 YEARS REMAINING. A HEX DUMP IS TO BE PERFORMED AT THE NEXT FOLLOW UP VISIT TO DETERMINE AN ACCURATE LONGEVITY ESTIMATE. TO DATE, THERE HAVE BEEN NO ADVERSE PATIENT EFFECTS REPORTED.

Description of Event or Problem · 1

BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THIS DEVICE WAS EXPLANTED AND REPLACED DUE TO BATTERY DEPLETION. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 INSIGNIA IMPLANTABLE PULSE GENERATOR NVZ GUIDANT CRM CLONMEL IRELAND 1291

Patients

Seq Age Sex Outcome Treatment
1 85 YR Hospitalization| L| R 4470| 4469| 1291