INSIGNIA
Report
- Report Number
- 2124215-2010-12816
- Event Type
- Injury
- Date Received
- July 27, 2010
- Date of Event
- October 9, 2009
- Report Date
- June 14, 2010
- Manufacturer
- GUIDANT CRM CLONMEL IRELAND
- Product Code
- NVZ
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
UPON RECEIPT IN OUR POST MARKET QUALITY ASSURANCE LABORATORY, A THOROUGH DEVICE EVALUATION WAS PERFORMED. THE DEVICE AT INCOMING, DISPLAYED IN MEMORY EVIDENCE THAT THE ADDRESS WAS CORRUPTED. THIS VALUE WAS CLEARED DURING TESTING. EXPOSURE TO RADIATION IS BELIEVED TO HAVE SET THE MEMORY LOCATION TO AN INCORRECT STATE. THE DEVICE WAS PUT THROUGH AND PASSED A SERIES OF AUTOMATED DIAGNOSTIC TESTING THAT VERIFIED THE PERFORMANCE OF PACING, SENSING, AND RECORDING FUNCTIONS OF THE DEVICE COMMENSURATE WITH BATTERY VOLTAGE. ADDITIONAL TESTING WAS PERFORMED TO ADDRESS THE OBSERVED HIGH RATE PACING IN THE CLINICAL SETTING. RATE RESPONSE (XL) WAS PROGRAMMED ON, THE DEVICE WAS THEN SHAKEN AND THE PACING RATE INCREASED WHICH IS AS EXPECTED FOR HAVE THE XL ON. THE DEVICE THEN RETURNED TO A NORMAL PACING RATE AT REST. NO OTHER ADVERSE EVENTS HAVE BEEN REPORTED. THIS ISSUE WILL BE RE-EVALUATED IF ADDITIONAL INFORMATION IS RECEIVED.
BOSTON SCIENTIFIC CRM RECEIVED INFORMATION THAT THIS PACEMAKER DEPENDENT PATIENT WAS RECEIVING RADIATION TREATMENT FOR PROSTATE CANCER. THE PATIENT CAME INTO THE CLINIC AFTER A TREATMENT AND PACING AT THE MAXIMUM SENSOR RATE (MSR) OF 130PPM WAS OBSERVED. THE CLINICIAN TURNED OFF RATE RESPONSE AND PACING RESUMED AT THE LOWER RATE LIMIT (LRL). HOWEVER, WHEN RATE RESPONSE WAS TURNED BACK ON, MSR PACING BEGAN WITHIN A COUPLE OF MINUTES. THE PACEMAKER WAS EXPLANTED AND REPLACED WITHOUT FURTHER INCIDENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | INSIGNIA | IMPLANTABLE PULSE GENERATOR | NVZ | GUIDANT CRM CLONMEL IRELAND | 1291 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 74 YR | MISMATCH| 4456| 1291 |