FDA Adverse Event Malfunction Summary report: N

VITALITY 2

MDR report key: 1771367 · Received July 27, 2010

Report

Report Number
2124215-2010-12494
Event Type
Malfunction
Date Received
July 27, 2010
Date of Event
June 14, 2010
Report Date
June 22, 2010
Manufacturer
GUIDANT CRM CLONMEL IRELAND
Product Code
LWS
Removal / Correction Number
Z-1047-2007 TO Z-1055-20
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

INITIAL ANALYSIS DISCOVERED THAT THE DEVICE'S BATTERY CAPACITY WAS LESS THAN EXPECTED. THIS EVENT WILL BE UPDATED AS ADDITIONAL INFORMATION BECOMES AVAILABLE.

Additional Manufacturer Narrative · 1

ANALYSIS WAS PERFORMED AT OUR POST MARKET QUALITY ASSURANCE LABORATORY. PORTIONS OF THE CIRCUITRY, INCLUDING THE BATTERY, WERE ISOLATED AND TESTED. FINAL ANALYSIS CONCLUDED THAT THIS DEVICE HAD TWO COMPROMISED LOW-VOLTAGE CAPACITORS, WHICH RESULTED IN A HIGH CURRENT CONDITION. THIS ISSUE IS DISCUSSED IN THE Q2 2010 PRODUCT PERFORMANCE REPORT. THIS EVENT WILL BE UPDATED IF ANY ADDITIONAL INFORMATION BECOMES AVAILABLE.

Description of Event or Problem · 1

BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THIS IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) WAS EXPLANTED DUE TO NORMAL BATTERY DEPLETION, PER PROCEDURE PAPERWORK. A NEW BOSTON SCIENTIFIC DEVICE WAS SUCCESSFULLY IMPLANTED, WITH NO ADVERSE PATIENT EFFECTS REPORTED. THIS EXPLANTED DEVICE WAS RETURNED FOR ANALYSIS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 VITALITY 2 IMPLANTABLE CARDIOVERTER DEFIBRILLATOR LWS GUIDANT CRM CLONMEL IRELAND T175

Patients

Seq Age Sex Outcome Treatment
1 50 YR T175| 0185| E102