VITALITY 2
Report
- Report Number
- 2124215-2010-12418
- Event Type
- Malfunction
- Date Received
- July 27, 2010
- Date of Event
- June 10, 2010
- Report Date
- July 9, 2010
- Manufacturer
- GUIDANT CRM CLONMEL IRELAND
- Product Code
- LWS
- Removal / Correction Number
- Z-1047-2007 TO Z-1055-20
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- PHYSICIAN
Narratives
UPON RECEIPT TO THE POST MARKET QUALITY ASSURANCE LABORATORY, ANALYSIS WILL BE PERFORMED IN AN ATTEMPT TO DETERMINE THE ROOT CAUSE OF THIS EVENT.
UPON RECEIPT AT OUR POST MARKET QUALITY ASSURANCE LABORATORY, DEVICE INTERROGATION REVEALED THAT THE DEVICE BATTERY STATUS WAS ELECTIVE REPLACEMENT INDICATOR (ERI). A LONGEVITY CALCULATION CONFIRMED THAT THE DEVICE DID NOT MEET LONGEVITY ESTIMATES PROVIDED IN DEVICE LABELING. PORTIONS OF THE CIRCUITRY, INCLUDING THE BATTERY, WERE ISOLATED AND TESTED. ALL LOW VOLTAGE POWER SUPPLY MEASUREMENTS WERE CURRENTLY WITHIN AN APPROPRIATE RANGE; HOWEVER, FINAL ANALYSIS CONCLUDED THAT THE PREMATURE BATTERY DEPLETION WAS DUE TO TWO COMPROMISED LOW VOLTAGE CAPACITORS, WHICH RESULTED IN A HIGH CURRENT CONDITION.
BOSTON SCIENTIFIC CRM RECEIVED INFORMATION THAT THIS DEVICE DECLARED ELECTIVE REPLACEMENT INDICATORS AFTER FIFTY-TWO MONTHS OF IMPLANT. THERE WAS CONCERN THAT THE BATTERY HAD DEPLETED MORE RAPIDLY THAN EXPECTED. A REPLACEMENT PROCEDURE WAS PERFORMED; THIS DEVICE WAS REMOVED, REPLACED AND RETURNED FOR ANALYSIS. NO ADVERSE PATIENT EFFECTS WERE REPORTED.
- -
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | VITALITY 2 | IMPLANTABLE CARDIOVERTER DEFIBRILLATOR | LWS | GUIDANT CRM CLONMEL IRELAND | T165 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |