FDA Adverse Event Malfunction Summary report: N

COMMUNICATOR

MDR report key: 1771257 · Received July 27, 2010

Report

Report Number
2124215-2010-12372
Event Type
Malfunction
Date Received
July 27, 2010
Date of Event
May 20, 2010
Report Date
June 7, 2010
Manufacturer
EXTERNAL MANUFACTURER
Product Code
LWS
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE COMMUNICATOR WAS RETURNED FOR ANALYSIS. THE POST MARKET QUALITY ASSURANCE LABORATORY CONFIRMED THE REPORTED ALLEGATION. THE COMMUNICATOR FAILED TO POWER UP WITH THE RETURNED POWER CORD. THE POWER CORD DID NOT BECOME HOT TO THE TOUCH DURING ANALYSIS, HOWEVER, THE CASE ON THE POWER CORD WAS OBSERVED TO BE MELTED.

Description of Event or Problem · 1

BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THE POWER CORD FOR THE COMMUNICATOR BECAME HOT TO THE TOUCH. NO PATIENT INJURY WAS REPORTED. A REPLACEMENT COMMUNICATOR WAS SHIPPED TO THE PATIENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 COMMUNICATOR ACCESSORY LWS EXTERNAL MANUFACTURER 6476

Patients

Seq Age Sex Outcome Treatment
1 82 YR 6024| E102| 0125| 1860| 1821