FDA Adverse Event
Malfunction
Summary report: N
COMMUNICATOR
MDR report key: 1771257
·
Received July 27, 2010
Report
- Report Number
- 2124215-2010-12372
- Event Type
- Malfunction
- Date Received
- July 27, 2010
- Date of Event
- May 20, 2010
- Report Date
- June 7, 2010
- Manufacturer
- EXTERNAL MANUFACTURER
- Product Code
- LWS
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
THE COMMUNICATOR WAS RETURNED FOR ANALYSIS. THE POST MARKET QUALITY ASSURANCE LABORATORY CONFIRMED THE REPORTED ALLEGATION. THE COMMUNICATOR FAILED TO POWER UP WITH THE RETURNED POWER CORD. THE POWER CORD DID NOT BECOME HOT TO THE TOUCH DURING ANALYSIS, HOWEVER, THE CASE ON THE POWER CORD WAS OBSERVED TO BE MELTED.
Description of Event or Problem · 1
BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THE POWER CORD FOR THE COMMUNICATOR BECAME HOT TO THE TOUCH. NO PATIENT INJURY WAS REPORTED. A REPLACEMENT COMMUNICATOR WAS SHIPPED TO THE PATIENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | COMMUNICATOR | ACCESSORY | LWS | EXTERNAL MANUFACTURER | 6476 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 82 YR | 6024| E102| 0125| 1860| 1821 |