VITALITY 2
Report
- Report Number
- 2124215-2010-12127
- Event Type
- Malfunction
- Date Received
- July 27, 2010
- Date of Event
- June 4, 2010
- Report Date
- June 8, 2010
- Manufacturer
- GUIDANT CRM CLONMEL IRELAND
- Product Code
- LWS
- Removal / Correction Number
- Z-1047-2007 TO Z-1055-20
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER
Narratives
UPON RETURN TO BOSTON SCIENTIFIC CRM'S POST-MARKET QUALITY ASSURANCE LABORATORY, A REVIEW OF DEVICE MEMORY REVEALED THE DEVICE DECLARED ITS ELECTIVE REPLACEMENT INDICATOR (ERI) AFTER EXPERIENCING TWO CONSECUTIVE CHARGE TIMES GREATER THAN THE EXTENDED MID-LIFE ERI CHARGE TIME LIMIT. ADDITIONAL LAB ANALYSIS AND TESTING SHOWED ERI WAS TRIGGERED EARLIER THAN EXPECTED BY EXTENDED CHARGE TIMES DUE TO A HIGHER-THAN-TYPICAL BUILD-UP OF DEVICE INTERNAL BATTERY IMPEDANCE. THE EOL (END OF LIFE) INDICATOR WAS NOT DECLARED WHILE THE DEVICE WAS IN SERVICE. TESTING CONFIRMED ALL DEVICE THERAPIES WERE AVAILABLE. THIS ISSUE IS DISCUSSED IN THE QUARTER 2, 2010 VERSION OF OUR PRODUCT PERFORMANCE REPORT.
UPON RECEIPT AT OUR POST MARKET QUALITY ASSURANCE LABORATORY, THIS DEVICE'S PROGRAMMED PARAMETER SETTINGS AND THERAPY HISTORY WERE USED TO ESTIMATE EXPECTED BATTERY USE TO DATE, THEN COMPARED TO ACTUAL USE. OUR INITIAL ANALYSIS SHOWED THIS DEVICE DID NOT MEET LONGEVITY EXPECTATIONS PER PROGRAMMED VALUES. ADDITIONAL ANALYSIS IS PENDING.
BOSTON SCIENTIFIC CRM RECEIVED INFORMATION THAT THIS DEVICE WAS EXPLANTED FOR NORMAL BATTERY DEPLETION. THERE WERE NO KNOWN ALLEGATIONS AGAINST THE DEVICE, AND NO REPORTS OF ADVERSE PATIENT EFFECTS. THIS DEVICE IS INCLUDED IN THE 4/5/2007 SHORTENED REPLACEMENT WINDOW ADVISORY POPULATION.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | VITALITY 2 | IMPLANTABLE CARDIOVERTER DEFIBRILLATOR | LWS | GUIDANT CRM CLONMEL IRELAND | T175 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 71 YR | T175| 0185 |