VITALITY 2
Report
- Report Number
- 2124215-2010-12240
- Event Type
- Injury
- Date Received
- July 27, 2010
- Date of Event
- June 7, 2010
- Report Date
- August 17, 2010
- Manufacturer
- GUIDANT CRM CLONMEL IRELAND
- Product Code
- LWS
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER
Narratives
AT THIS TIME, THE DEVICE HAS NOT BEEN RETURNED TO BOSTON SCIENTIFIC, CRM FOR ANALYSIS. THERE IS NO ADDITIONAL INFORMATION AVAILABLE. IF FURTHER INFORMATION BECOMES AVAILABLE, THIS EVENT WILL BE REOPENED.
ANALYSIS OF THE RETURNED PRODUCT IS CURRENTLY ONGOING. THIS EVENT WILL BE UPDATED UPON COMPLETION OF THE ANALYSIS.
A REVIEW OF DEVICE MEMORY REVEALED THAT THE DEVICE DECLARED ELECTIVE REPLACEMENT INDICATOR (ERI) AFTER EXPERIENCING TWO CONSECUTIVE CHARGE TIMES GREATER THAN THE EXTENDED MID-LIFE ERI CHARGE TIME LIMIT. THE OBSERVED RATE OF BATTERY USAGE WAS COMPARED TO THE EXPECTED RATE OF BATTERY USAGE; THE RESULTS INDICATED THAT THE MONITORING VOLTAGE WAS NORMAL. ALTHOUGH THE BATTERY ITSELF HAD NOT DEPLETED PREMATURELY, ERI WAS TRIGGERED EARLIER THAN EXPECTED BY EXTENDED CHARGE TIMES DUE TO A HIGHER-THAN-TYPICAL BUILD-UP OF INTERNAL BATTERY IMPEDANCE. THIS DEVICE IS INCLUDED IN THE MID-LIFE DISPLAY OF REPLACEMENT INDICATORS (NOVEMBER 27, 2007) ADVISORY POPULATION. THIS ISSUE IS DISCUSSED IN THE Q3 2010 PRODUCT PERFORMANCE REPORT.
BOSTON SCIENTIFIC CRM RECEIVED INFORMATION THAT THIS IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) WHICH WAS INCLUDED IN THE MID-LIFE DISPLAY OF REPLACEMENT INDICATORS PRODUCT UPDATE CLASSIFIED AS A PRODUCT ADVISORY WAS INITIALLY COMMUNICATED ON (B)(6) 2007, DECLARED END OF LIFE (EOL) DUE TO A 32.3 SECOND CHARGE TIME. THE MONITORING VOLTAGE WAS 2.68 VOLTS. THE DEVICE WAS EXPLANTED AND SUCCESSFULLY REPLACED. TO DATE, THERE HAVE BEEN NO ADVERSE PATIENT EFFECTS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | VITALITY 2 | IMPLANTABLE CARDIOVERTER DEFIBRILLATOR | LWS | GUIDANT CRM CLONMEL IRELAND | T167 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 71 YR | Required Intervention | T167| 0184| 4470 |