FDA Adverse Event Malfunction Summary report: N

CONTAK RENEWAL

MDR report key: 1771146 · Received July 27, 2010

Report

Report Number
2124215-2010-12139
Event Type
Malfunction
Date Received
July 27, 2010
Date of Event
June 7, 2010
Report Date
June 15, 2010
Manufacturer
GUIDANT CRM CLONMEL IRELAND
Product Code
NIK
Removal / Correction Number
Z-1192-06 THRU Z-1194-06
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

ANALYSIS OF THE RETURNED PRODUCT IS CURRENTLY IN PROGRESS. NO ADDITIONAL INFORMATION IS AVAILABLE. SHOULD ANY ADDITIONAL INFORMATION RELATED TO THIS EVENT BECOME AVAILABLE, THIS EVENT WILL BE UPDATED.

Additional Manufacturer Narrative · 1

ANALYSIS WAS PERFORMED AT OUR POST MARKET QUALITY ASSURANCE LABORATORY. PORTIONS OF THE CIRCUITRY, INCLUDING THE BATTERY, WERE ISOLATED AND TESTED. FINAL ANALYSIS CONCLUDED THAT THIS DEVICE HAD TWO COMPROMISED LOW-VOLTAGE CAPACITORS, WHICH RESULTED IN A HIGH CURRENT CONDITION. THIS ISSUE IS DISCUSSED IN THE Q2 2010 PRODUCT PERFORMANCE REPORT.

Description of Event or Problem · 1

BOSTON SCIENTIFIC CRM RECEIVED INFORMATION THAT THIS CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATOR (CRT-D) WAS EXPLANTED DUE TO NORMAL BATTERY DEPLETION. A BOSTON SCIENTIFIC IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) WAS SUCCESSFULLY IMPLANTED. THIS DEVICE IS PART OF THE SHORTENED REPLACEMENT WINDOW ADVISORY. THIS ADVISORY WAS ORIGINALLY COMMUNICATED (B)(6) 2007. UPON RECEIPT, INITIAL ANALYSIS DETERMINED THAT THE DEVICE DID NOT MEET EXPECTED LONGEVITY PREDICTIONS AS DESCRIBED IN LABELING. NO ADVERSE PATIENT EFFECTS HAVE BEEN REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CONTAK RENEWAL IMPLANTABLE CHF GENERATOR NIK GUIDANT CRM CLONMEL IRELAND H179

Patients

Seq Age Sex Outcome Treatment
1 66 YR