FDA Adverse Event
Malfunction
Summary report: N
UNICEL DXI 600 SYNCHRON CHEMISTRY ANALYZER
MDR report key: 1771125
·
Received July 27, 2010
Report
- Report Number
- 2122870-2010-00320
- Event Type
- Malfunction
- Date Received
- July 27, 2010
- Date of Event
- June 12, 2010
- Report Date
- July 23, 2010
- Manufacturer
- BECKMAN COULTER INC.
- Product Code
- JJE
- PMA / PMN Number
- K023764
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JO
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
A BCI FIELD SERVICE ENGINEER (FSE) PERFORMED DIAGNOSTIC TEST; ALL TESTS PASSED EXCEPT THE CARRYOVER TESTS. FSE REPLACED PARTS AND REPEATED THE CARRYOVER TEST. THE CUSTOMER SENT THE SAMPLE TO BECKMAN CUSTOMER PRODUCT LINE SUPPORT (CPLS) FOR ADDITIONAL TESTING. CPLS TESTING CONFIRMED A PATIENT SOURCE INTERFERENCE WITH A SAMPLE FROM THIS PATIENT.
Description of Event or Problem · 1
A CUSTOMER CONTACTED BECKMAN COULTER INC. (BCI) IN REGARDS VITAMIN B12 RESULT ABOVE THE NORMAL REFERENCE RANGE UNICEL DXL 600 ACCESS 2 IMMUNOASSAY ANALYZER.PATIENT SAMPLE WAS REPORTED OUT OF THE LABORATORY.THE SAMPLE WAS REPEATED ON AN ALTERNATE METHOD AND LOWER RESULT WAS OBTAINED. PATIENT TREATMENT WAS NOT IMPACTED BY THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | UNICEL DXI 600 SYNCHRON CHEMISTRY ANALYZER | CLINICAL CHEMISTRY ANALYZER | JJE | BECKMAN COULTER INC. |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |