FDA Adverse Event Malfunction Summary report: N

UV-FLASH SOLUTION TRANSFER SET(SHORT) - JAPAN

MDR report key: 1771041 · Received July 27, 2010

Report

Report Number
1423500-2010-01656
Event Type
Malfunction
Date Received
July 27, 2010
Date of Event
July 1, 2010
Report Date
July 6, 2010
Manufacturer
BAXTER HEALTHCARE - MOUNTAIN HOME
Product Code
KDJ
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE SAMPLE HAS BEEN RECEIVED FOR EVALUATION. A FOLLOW-UP REPORT WILL BE SUBMITTED UPON COMPLETION OF THE EVALUATION. A 510(K) NUMBER WILL NOT BE PROVIDED IN THE EMDR AS THIS PRODUCT IS MANUFACTURED FOR DISTRIBUTION OUTSIDE OF THE U.S. AND DOES NOT HAVE A 510(K) NUMBER. HOWEVER, IT IS BEING REPORTED BECAUSE IT IS THE SAME AS OR SIMILAR TO PRODUCT DISTRIBUTED WITHIN THE U.S.

Additional Manufacturer Narrative · 1

(B)(4). THE CUSTOMER REPORT OF DIFFICULTY USING THE TRANSFER SET WAS NOT CONFIRMED IN THE LAB. DURING VISUAL INSPECTION NO MANUFACTURING ABNORMALITIES WERE NOTED. A LAB TITANIUM ADAPATER WAS FUNCTIONALLY HAND TIGHTENED WITHOUT DIFFICULTY. THE SET WAS TESTED UNDERWATER AT 8 PSI WITH NO LEAK NOTED. THE SLEEVE OF THE SET WAS OPENED/CLOSED SEVERAL TIMES WITHOUT DIFFICULTY WITH THE SLEEVE FITTING IN THE INDENT PROPERLY. A BATCH REVIEW WAS NOT PERFORMED AS THE LOT NUMBER WAS UNKNOWN. THE ROOT CAUSE WAS UNDETERMINED. RENAL QUALITY ENGINEERING, ALONG WITH PLANT QUALITY AND MANUFACTURING PERSONNEL, WILL CONTINUE TO MONITOR THIS PRODUCT LINE FOR TRENDS AND WILL TAKE CORRECTIVE/PREVENTIVE ACTION AS APPROPRIATE.

Description of Event or Problem · 1

BAXTER (B)(4) REPORTED THE CLAMP ON THE TRANSFER SET WAS LOOSE TO CLOSE. AIR BUBBLES WERE OBSERVED IN THE TUBING WHEN THE SLEEVE WAS OPENED. THERE WAS NO PATIENT INJURY OR MEDICAL INTERVENTION. THE ACTUAL SAMPLE IS AVAILABLE FOR EVALUATION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 UV-FLASH SOLUTION TRANSFER SET(SHORT) - JAPAN SET, ADMINISTRATION, FOR PERITONEAL DIALYSIS, DISPOSABLE KDJ BAXTER HEALTHCARE - MOUNTAIN HOME UNKNOWN

Patients

Seq Age Sex Outcome Treatment
1