Report

Report Number
3004034546-2023-00001
Event Type
Injury
Date Received
September 8, 2023
Report Date
September 8, 2023
Manufacturer
A & I INDUSTRIES LTD
Product Code
ITJ
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 0

WE DID THE FINAL QUALITY INSPECTION BEFORE DELIVERY. THE ROLLATOR MODEL PASSED THE STATIC LOAD TEST & FATIGUE TEST THAT WERE PERFORMED IN 2023/05 ACCORDING TO THE ISO 11199-2:2021.

Description of Event or Problem · 0

DRIVE DEVILBISS HEALTHCARE WAS NOTIFIED OF AN INCIDENT INVOLVING A ROLLATOR BY AN END USER'S DAUGHTER, WHO STATED THAT THE "RIGHT HANDLE SUDDENLY BENT DOWN WHERE THE PLASTIC HANDLE MEETS THE METAL FRAME" AND REPORTED THAT HER MOTHER LOST HER BALANCE AND BROKE HER LEG. THE END USER WAS TAKEN TO THE HOSPITAL AND UNDERWENT A WEEK OF TREATMENT FOR HER INJURIES. DRIVE REQUESTED THE UNIT BE RETURNED FOR INVESTIGATION AND WILL FILE AN UPDATE IF ADDITIONAL INFORMATION BECOMES AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1501432 DRIVE ROLLATOR ITJ A & I INDUSTRIES LTD RTL10266

Patients

Seq Age Sex Outcome Treatment
1 Female Hospitalization